User Manual
Table Of Contents
- Manufacturer
- Identification and Publication Details
- Regulatory
- Explanation of Symbols
- Conventions
- Important Information
- About
- Intended Use
- Compatibility
- Indications for Use
- Contra-indications
- Training
- Safety
- Electromagnetic Compatibility (EMC)
- Using Batteries Safely
- Examining the Shipment
- Initial Setup
- Rear Panel Connections
- Understanding Battery Operations
- Understanding Wireless Network Operations
- Advanced User Options
- Accessory List
- System Overview
- Getting Started
- Alarms
- Monitoring ECG
- ECG Monitoring Considerations for the MR Environment
- wECG Module and ECG Lead Cable
- Quadtrode Electrodes
- Work Flow for ECG Monitoring
- Selecting the ECG Lead Cable and Quadtrode Electrode Type
- Identifying the Placement Site for the Quadtrode Electrode
- Preparing the Quadtrode Electrode Site
- Attaching the ECG Lead Cable
- Checking the ECG Signal Strength
- Minimizing ECG Waveform Noise
- Positioning the ECG Lead Cable and wECG Module for Scanning
- ECG Waveforms and VS Box
- ECG Menu
- Monitoring SPO2
- Monitoring CO2 (LoFlo Option)
- Monitoring Invasive Blood Pressure
- Indications and Contraindications
- Patient Preparation for IBP Monitoring
- Transducer Component, Connection, and Feature Locations
- MR 400 Preparation for IBP Monitoring
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- I. Connecting the Reusable Cable to the MR400
- II. Kit Set Up
- III. Purging Air from the Monitoring Line
- IV. Zeroing, Leveling and Calibration
- V. Connecting the Monitoring Kit to the Patient
- The IBP transducer must not be mounted to the patient, or patient burn may result.
- VI. Fast Flushing
- VII. Checking for Leaks
- VIII. In the MR Room
- Neonatal Patients: Expression MR IBP DPT Kit, I/N (REF 989803194641)
- Adult and Pediatric Patients: Expression MR IBP DPT Kit, A/P (REF 989803194631)
- Zeroing the Pressure Transducer
- P1 (and P2) Menu
- Monitoring Agents and Gases (AGENT Option)
- Monitoring RESP
- Monitoring Temperature
- Monitoring NIBP
- Trend Data and Printing
- Maintenance and Troubleshooting
- General Cleaning Guidelines
- Removing all Power to the MR400
- Removing Power from the Wireless Modules
- User Routine-Checks and Planned Maintenance
- Cleaning, Disinfection, and Damage Inspection
- Sterilization
- Testing Alarms
- Testing a Dropped Wireless Module
- Verification Testing
- Anesthetic Oxygen (O2) Sensor Depletion
- Updating Software
- Calibrating the Touch Screen
- Troubleshooting
- Repair
- Environmental Requirements
- Passing the Product on to another User
- Final Disposal of the Product
- Specifications
- Warranty
- Regulatory Information
- Gating Feature
- Guidelines and References
ExpressionMR400InstructionsforUseImportantInformation1‐21
Understanding Battery Operations
Cart Batteries
Warning
WARNING
Do not touch the patient and the circuitry in the battery compartments of the MR400
simultaneously.
Cartbatteries,wheninstalled(main)andswitchedon(reserve),arechargedand c onditionedby
anintegratedchargingsystem.WhenturnedonandconnectedtoACmains,theMR400operates
fromACpowerandsimultaneouslychargesallcartbatteries.Whenturnedoffandconnectedto
ACmains,batterychargingfunctionscontin ue.
Ifatanytime,ACmainsislost,theMR400willautomaticallyswitchtobatterypowertoprovide
uninterruptedservice—then,whenACmainsisrestored,theMR400willautomatically,without
delay,revertbacktoACpowerfunctions.IfthereservebatteriesarefullydepletedandACmains
islost,thenthe
unitwillpoweroff.
TheMR400’smaximumoperatingtimeonbatterypowerdependsupontheenabledparameters
andthetypeandfrequencyofmonitoringfunctions(seetheBattery,OperationTimeonpageA‐4
foralisting).
Charging Cart Batteries
Cartbatteriesmustbechargedbeforeinitialuse.Duringinitialsetuporwheninstallingnew
batteries,chargethecartbatteriesforatleast12hourssothattheyarefullychargedand
conditionedforoperation.
Notes
• Use only with the specified battery charger.
• The main batteries must always be inserted and the reserve batteries must always be
switched on to prevent loss of patient monitoring during a power outage. If main batteries
are not inserted and if the reserve batteries are not switched on, then during power outage
unsaved user settings will revert to factory defaults.
• We recommend plugging the MR400 into a backup generator or equivalent means to
prevent a lapse in patient monitoring during a power loss.