User's Manual

A-4
NON-INVASIVE BLOOD PRESSURE
GENERAL Oscillometric method (with inflatable cuff).
Determines systolic, diastolic and mean arterial
pressures, and pulse rate.
Patient Types Adult and Neonate.
PNEUMATIC SYSTEM
Cuff Inflation Pressure Initially 170 mmHg for Adult/Pediatric. (120
mmHg for Neonate). Subsequent inflation
pressures determined by last measured systolic
pressure.
Cuff Inflation Pressure Range Adult: 110 to 265 mmHg.
Neonate: 80 to 135 mmHg.
Overpressure Protection Automatically releases cuff pressure if inflation
pressure exceeds 285 mmHg (adult, for
neonate: 150 absolute or greater than 142 but
less than 150 for 15 seconds).
MEASUREMENT RANGE
Systolic 25 to 260 mmHg.
Diastolic 10 to 235 mmHg.
Mean Arterial 10 to 255 mmHg.
(Measurements are possible only in pulse rate range of 40 to 200 bpm for adult and 40 to 230
bpm for neonate.)
ACCURACY
Pulse Rate 2% full scale.
Pressure Zero Offset 20 mmHg, ±5 mmHg.
Pressure Span Accuracy ±3 mmHg.
Pressure Transducer Range 0 to 307 mmHg.
ALARM LIMITS
Systolic, Mean and Diastolic Minimum: 5 to 249 mmHg.
Maximum: 5 to 249 mmHg.
Pulse (when "HR" derived from NIBP) Minimum: 30 to 249 bpm.
Maximum: 60 to 249 bpm.
MODES
Manual Immediate upon operator command.
Automatic Determinations automatically made with
selectable intervals of 1, 2, 2.5, 3, 5, 10, 15, 20,
30 and 45 minutes, and 1, 2 and 4 hours.
STAT Up to five (5) consecutive measurements (five
(5) minute maximum time duration).