User's Manual

3-12
The following is a description of the PARAMETERS SELECTION Menu
options:
(a) ECG. Selecting this menu option will turn the ECG display ON
(default) or OFF. The heart rate will remain on the screen, allowing it to
be displayed from another source, if the heart-rate source (the HR
SOURCE selection) is set to AUTO.
(b) NIBP. Selecting this menu option switches the NIBP ON (default) and
OFF.
(c) P1. This option is not available on the 3160 MRI Physiological
Monitoring System.
(d) P2. This option is not available on the 3160 MRI Physiological
Monitoring System.
(e) P3. This option is not available on the 3160 MRI Physiological
Monitoring System.
(f) P4. This option is not available on the 3160 MRI Physiological
Monitoring System.
(g) SPO2. Selecting this menu option switches SpO2 ON and OFF.
(h) EtCO2. Selecting this menu option switches EtCO2 ON and OFF.
(i) RESP. This option is not available on the 3160 MRI Physiological
Monitoring System.
(j) TEMP. This option is not available on the 3160 MRI Physiological
Monitoring System.
(k) AUX. This option is not available on the 3160 MRI Physiological
Monitoring System.
(l) AGENTS. (If option is installed) Selecting this menu option switches
the Anesthetic Agent Option ON and OFF (default).
(m) RETURN. Selecting this menu option returns the monitor to the
Normal Screen.
(2) Sound Adjust. Selecting this menu option will bring up the SOUND
ADJUST Menu (See Figure 3-12) which allows the user to switch the Alarm
Tone ON and OFF, set the heart-rate tone source and set the volume for the
different sounds the 3160 MRI Physiological Monitoring System produces.
While in this menu, all real tones are disabled and the message “REAL
TONES DISABLED” is displayed at the top of the screen. Note that only the
sound is disabled and the violated alarms will still flash on the screen if the
parameter's Alarm Limit is violated. This menu has a time-out feature. If no
action is taken for approximately 60 seconds, the monitor will automatically
return to the Normal Screen.
NOTE
The WPU does not contain a speaker and, therefore, it will not make a sound. The following settings
will affect the operation of the DCU.
WARNING
The Alarm Tone can be set to OFF. Always check that the Alarm Tone setting is appropriate for each
particular patient. Alarm Sound volume is adjustable for suitability to various clinical environments
(where background noise may range from relatively quiet to noisy). Always verify that the user/
attendant of this monitor can hear the Alarm Sound above the ambient noise.