(b) SIMULATION MODE. This menu option allows the operator to turn
the Simulation Mode ON. When selected the monitor will first display
a YES/NO Menu and require user confirmation before entering the
Simulation Mode. While in the Simulation Mode the displayed patient
information is computer generated and not actual patient
determinations. As a safety feature, while in the Simulation Mode the
message “SIMULATION” is displayed in the center of the screen and,
when printing any strip or chart, “SIMULATION” will appear on the
printout. To exit the Simulation Mode, the monitor must be powered
Off.
(c) NIBP TESTS. Selecting this menu option will bring up the NIBP
TESTS Menu (See Figure 3-15).
The following options are provided in this menu:
CALIBRATE. Selecting this menu option will display
NIBP CAL Offset Pressure and actual Pressure Reading in
the NIBP TESTS menu. These are used to verify and
calibrate the internal NIBP.
Figure 3-15. NIBP TESTS Menu
LEAK TEST. Selecting this menu option will display
NIBP LEAK TEST with the Peak (beginning) Pressure
and Final (current) Pressure displayed, along with the
number of Passes and Failures of the test, to determine the
leak rate of the NIBP system. To begin this test, highlight
the LEAK TEST menu option and press the Rotary Knob;
to stop a test in progress, press the Rotary Knob a second
time.
WARNING
The Simulation Mode will display real looking waveforms which are computer generated. The
monitor will not monitor patients while in the Simulation Mode. Do not activate the Simulation
Mode when this monitor is connected to a patient. To exit the Simulation Mode, the monitor must
be powered Off.
WARNING
The Leak Test feature is for use by qualified service personnel only.
Never initiate a Leak Test while the cuff is applied to a patient. Continuous cuff pressure could lead
to patient injury.