User's Manual

4-25
4.5.8
Oxygen Monitoring. This parameter is part of the EtCO2/Agents module. That is, it uses the
exhaust air from the EtCO2/Agents module as its imput and its air flow is controlled by the pneumatic
hardware in the EtCO2/Agents hardware module. The O2 value is displayed in the Anesthetic Agents
Display (See Figure 4-17).
a. Oxygen Module Calibration. The O2 module is calibrated automatically upon
power-up (as described in the NOTE above). Thereafter, is calibrated everytime an O2
CAL is initiated by the user (See the GAS CAL menu (SETUPS/SERVICE(BIO-
MED)/GAS CAL) for further information). The operator is prompted that an O2
calibration is required if O2 remains above 22% for 12 hours and, since O2 calibration
requires one (1) minute, the operator may reject the calibration request by answering
NO to the YES/NO prompt. If NO is selected the operator is then prompted every 30
minutes until the O2 calibration is initiated by a YES response. Calibration is done
based on the assumption that the ambient O2 concentration, sensed at the EtCO2
Zeroing Port, is 20.9% (during calibration, the patient GAS Input Port must not be
connected to an oxygen source).
b. Oxygen Sensor Replacement. The Oxygen Sensor (Invivo Part Number 9445) is
located on the rear of the WPU near the DCU mounting pole. The Oxygen Sensor has
an expected life of greater than six months with expected life inversely proportional to
changes in Oxygen Concentration, Temperature and Pressure. The Oxygen Sensor
begins aging immediately upon the opening of the package and should, therefore, not
be opened until ready for use. The Oxygen Sensor should be replaced periodically as
part of routine maintenance. This monitor cannot be connected to the scavenge gas
system without the Oxygen Sensor installed, never operate without the monitor
connected to the scavenge gas system. Always operate the monitor with the Oxygen
Sensor installed.
(1) Oxygen Sensor Replacement Instructions. Replace the Oxygen Sensor as
follows:
(a) Unplug the adapter cable from the Oxygen Sensor.
(b) Unscrew the Oxygen Sensor by turning it counterclockwise until it
comes free.
(c) Open bag containing new disposable Oxygen Sensor (IRI #9445).
(d) Screw Oxygen Sensor into threaded port on Adaptor Subassembly.
(e) Plug adapter cable assembly into the Oxygen Sensor.
READJUSTING CO2
ZERO
Message flashing red. Occurs during CO2/Anesthesia
Gas Sensor zeroing operation.
REPLACE O2 SENSOR The Oxygen Sensor is bad and must be replaced.
NOTE
When the Agents parameter is turned on the O2 sensor is automatically calibrated during initial
warm up of the EtCO2/Agent Module. During this time the patient GAS input port must not be
connected to an oxygen source as the monitor requires 20.9% Oxygen (room air) for automatic
calibration.