5.1.1 EX3DV3 E-Field Probe Specification
Construction Symmetrical design with triangular core
Built-in optical fiber for surface detection
System (EX3DV3 only)
Built-in shielding against static charges
PEEK enclosure material
(resistant to organic solvents, e.q., glycol)
Calibration In air from 10 MHz to 6 GHz
In brain and muscle simulating tissue at
frequencies of 900MHz, 1800MHz, 5200MHz
and 5500MHz and 5800MHz (accuracy ±8%)
Calibration for other liquids and frequencies upon request
Frequency 10 MHz to > 6 GHz; Linearity: ±0.2 dB
(30 MHz to 3 GHz)
Directivity ±0.3 dB in brain tissue (rotation around probe axis)
±0.5 dB in brain tissue (rotation normal probe axis)
Dynamic Range 10μW/g to > 100mW/g; Linearity: ±0.2dB
Surface Detection ±0.2 mm repeatability in air and clear liquids
over diffuse reflecting surface(EX3DV3 only)
Dimensions Overall length: 330mm
Tip length: 20mm
Body diameter: 12mm
Tip diameter: 2.5mm
Distance from probe tip to dipole centers: 1.0mm
Application General dosimetry up to 6GHz
Compliance tests of mobile phones
Fast automatic scanning in arbitrary phantoms
Figure 4.
EX3DV3 E-field Probe
Figure 5.
Probe setup on robot