User's Manual Part 3

8-11
Problem
Possible Cause
Solution
Aged battery will not
hold charge.
1) Plug AC Adapter
power cord into an
AC outlet
immediately.
2) Plug AC Adapter
into the circular
power receptacle on
the rear of the
device.
3) For 1133 batteries
only. After a few
hours of charge
check battery by
pressing the test
button on the
battery.
4) If problem persists
dispose of the
battery according to
policy.
5) Replace battery or
POD with a new
one.
6) If problem persists,
contact IRadimed
technical support.
8
Battery is hard to
remove.
Swollen Battery Cells.
Return the device to
IRadimed for battery
removal.
8.6.2. Troubleshooting Alarm Related Problems
Problem
Possible Cause
Solution
1
No Audible alarm tone
heard.
Alarm volume set too low
for the use environment.
1) Press the Settings
Button.
2) Select Alarms
Function
3) Adjust alarm volume
for the intended use
environment.
Faulty alarm speaker.
1) Contact IRadimed
Technical Support.
2
Continuous alarm tone
after alarm silence is
pushed.
Faulty hardware.
1) Note any onscreen
messages.
2) Contact IRadimed
technical support.