User's Manual Part 3

8-20
Problem
Possible Cause
Solution
5
Temperature error or
intermittent function
Temperature sensor is
not securely attached to
the patient.
Re-attach Temperature
sensor.
Poor Temperature sensor
placement
Reposition/replace
Temperature sensor.
Faulty Temperature
sensor
Replace Temperature
sensor, or refer to
qualified technical service
representative.
Temperature Vital Signs
are not performing as
expected
All other issues
1) Ensure the
Temperature
parameter is
enabled.
2) Inspect all
Temperature cable
for damage or
kinking and replace
as necessary.
3) Review section 6.4
of the operators
manual.
4) Contact your local
representative to
schedule
applications training.
6
Respiration and Gases
from 3886 are not
performing as expected
1) Ensure the 3886 is
turned ON
2) Ensure the Gas
parameter is
enabled on the
3880.
3) Inspect all EtCO2
components for
damage and leaks
and replace as
necessary.
4) Review section 6.3.8
of the operators
manual.
5) Contact your local
representative to
schedule
applications training.