User's Manual Part 3

10-4
10.1.1.6. Electrical Specifications
Technical Parameters
Technical Detail
Power
Requirements
Voltage Range:
(All 3880 system components)
85 - 264 VAC
Frequency Range:
50 - 60 Hz
Max Consumption:
3880 Monitor
3885-B Base Station
< 40 VA during charging
< 65 VA during charging, 3885-B
3886 Multi-Gas Unit
< 10 VA
Battery Capacity
3880 Monitor:
14.8 V at 6 Ah Lithium Polymer
3885-T Remote Tablet:
7.4 V at 6 Ah Lithium Polymer
3881/3882 W ireless PODS:
3.7 V at 1200 mAh Lithium Polymer
Battery Operation
Time
3880 Monitor:
>8 hours with NIBP readings every 5 minutes
3885-T Remote Tablet:
>10 hours
3881/3882 Wireless PODS:
>12 hours
Battery Charge Time
3880 Monitor:
< 5 hours to 90% capacity
3885-T Remote Tablet:
< 5 hours to 90% capacity
3881/3882 Wireless PODS:
< 3 hours to 90% capacity
Power On
Boot Time:
< 4 seconds
10.1.1.7. Environmental Specifications
Technical Parameters
Technical Detail
Operating
All 3880 system components
Temperature Range:
+10
to + 40
C (+50
to + 104
F)
Humidity Range:
5% to 85% RH, non-condensing
Altitude Range:
0 5000 meters, Pressure: 1010hPa 540hPa
Storage
All 3880 system components
Temperature Range:
-20
to + 50
C (-4
to + 122
F)
Humidity Range:
5% to 95% RH, non-condensing
Altitude Range:
0 5000 meters, Pressure 1010hPa 540hPa
10.1.1.8. MRI Conditions
Technical Parameters
Technical Detail
3880 Monitor
MR Environment Safety:
MRI Conditional
Magnetic Field Limit:
30,000 Gauss
MRI System:
0.5 to 3.0 Tesla MRI Systems
3881/3882 Wireless PODS
MR Environment Safety:
MRI Conditional
SAR:
4 W/kg whole body average SAR
Magnetic Field Limit:
30,000 Gauss
MRI System:
0.5 to 3.0 Tesla MRI Systems
3885-T Remote Tablet
MR Environment Safety:
MRI Conditional
Magnetic Field Limit:
15,000 Gauss
3885-B Base Station
MR Environment Safety:
MRI Unsafe
Accessories
MR Environment Safety:
MRI Safe as listed in Section 9.1-9.7
3886 Multi-Gas Unit
Magnetic Field Limit:
MR conditional 600 gauss