User's Manual Part 3

10-6
10.1.1.11.2. SpO
2
Technical Parameters
Technical Detail
Technique:
Masimo SET
®
Saturation Range:
1% - 100%
Saturation Accuracy:
± 2.1% at 70% - 99% (full scale)
< 70% oxygen saturation is unspecified
Saturation Resolution:
1%
Pulse Rate Range:
30 - 240 ppm
Pulse Rate Accuracy:
± 3 ppm
Pulse Rate Resolution:
1 pulse per minute (PPM)
Wavelength Range:
660 nm / 905 nm Note: Wavelength range can be especially useful to
clinicians
Emitted Light Energy
< 1.2mW maximum average at 905nm
Calibration Range:
70 - 100%
Minimum sensor Bend
Radius
4 cm (1.6 inches)
SpO2 averaging time:
6 seconds
10.1.1.11.3. NIBP
Technical Parameters
Technical Detail
Technique:
Oscillometric
Modes:
Manual, Automatic and STAT
Measurement Time:
< 60 seconds typical; standard adult cuff
Systolic Measureable
Pressure Range:
Adult/Pediatric: 40 - 270 mmHg (5.3 - 36 kPa)
Neonatal: 30 - 130 mmHg (4 - 17 kPa)
Diastolic Measureable
Pressure Range:
Adult/Pediatric: 25 - 245 mmHg (3.3 - 32 kPa)
Neonatal: 10 - 100 mmHg (1.3 - 13 kPa)
Mean Measureable
Pressure Range:
Adult/Pediatric: 30 - 255 mmHg (4 - 34 kPa)
Neonatal: 15 - 120 mmHg (2 - 16 kPa)
Pressure Accuracy:
Max. Std. Deviation: <8 mmHg (1.1 kPa)
Max. Mean Error: within ± 5mmHg (±0.7 kPa)
Pressure Resolution:
1 mmHg (0.1 kPa)
Pulse Rate Range:
30-240 ppm
Pulse Rate Accuracy:
± 1% or ± 5 BPM, whichever is greater
Max Cuff Inflation
Pressure:
Adult/Pediatric: 270 mmHg
Neonatal: 140 mmHg
Pressure Transducer
Range:
0 - 300 mmHg (0 - 40 kPa)
Transducer Accuracy:
The greater of ± 2 mmHg or 2% of the reading
Overpressure Protection:
Adult: 300 mmHg (40 kPa) < 2 seconds
Pediatric: 300 mmHg (40 kPa) < 2 seconds
Neonatal: 150 mmHg (20 kPa) < 2 seconds
Initial Pressure:
Adult: 165 mmHg (22 kPa)
Pediatric: 165 mmHg (22 kPa)
Neonatal: 100 mmHg (13.3 kPa)
All initial pressures ± 15 mmHg (2 kPa)
STAT Mode:
3 consecutive NIBP Readings
Minimum Time Between
Readings:
Auto: 30 seconds (non STAT)
Manual: 5 seconds