User's Manual Part 3

10-8
Desflurane
+12 % of reading @ 15 vol%
Ethanol
No effect at 0.3 vol%
Isopropanol
No effect at 0.5 vol%
Acetone / Metabolic Ketones
No effect at 1 vol%
Methane
No effect at 3 vol%
Carbon Monoxide
No effect at 1 vol%
Nitrogen Monoxide
No effect at 0.02 vol%
Oxygen
No effect at 100 vol%
10.1.1.11.7. O2 (Part of 3886)
Resolution
1%
Range
0 to 100 %
Accuracy 0 to 59%
+/- 3%
Accuracy 60 to 100%
+/- 5%
10.1.1.11.8. Temperature
Technical Parameters
Technical Detail
Technique:
Direct Fiber-Optic
Range:
30 - 44
C (86111.2
F)
Accuracy:
± 0.3
C (±0.54
F)
Extended Range:
10
C to 50
C (50
F to 122
F)
Extended Range
Accuracy:
±0.4
C (±0.72
F)
Resolution:
0.1
Response Time:
< 20 seconds
Application Type:
Axillary or skin surface
B. Messages, Alerts, Alarm Priority
Message
Trigger Condition
Priority
Apnea
Respiration detection system ( CO2 / Multi-Gas) reports Apnea
High
Agent Unit Connecting
Indicates 3880 attempting to make connection with 3886 Multi-gas Unit
N/A
Check Print Door
Printer door is not fully closed
N/A
CO2 Occlusion
Gas sampling line is occluded
Med
CO2 Overscale
Co2 measurement exceeds range of the display scale
Low
CO2 Zeroing
Performing a Zero reference in gas system
N/A
COMM LOSS
Wireless communication loss with 3885-B for more than 4 seconds
Low
Crit Mon Batt
5% of battery capacity in 3880 monitor
Med
Crit Tab Batt
5% of battery capacity in 3885-T Remote Tablet
Med
Data Delay
Display update error or delay possible
High
ECG Batt Crit
Battery level in ECG transmitter POD at extreme low level
Med
ECG Inop
Hardware or software failure detected
Med
ECG Lead Fail
Lead wire has no electrical connection to the patient or ECG level over-
scale
Low
EtCO2 Cal Error
Calibration error detected
Med