User's Manual

Flat Panel Detector Luna1013XE
58 User Manual of Luna1013XE
2. After receiving the Cmd_Prep, it will start the prepare process, and send back the acknowlage
of “Prohibit” and “Enable”, the “XWIN” will be started.
3. The max XWIN is configured by parameter “Exp Window Time” on “Detector” page
“Parameter” tab, the unit is “ms”.
4. User starts the X-Ray.
5. Send “SingleAcq” on UI “Acquire” page after the X-Ray is end.
6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.3. AED Mode
5.3.1. Inner
IDLE
XWIN1
(fixed)
X-Ray
imaging
XWIN1
(fixed)
Offset
imaging
X-Ray Info
RAW Preview
RAW X-Ray image
If HW-offset is disabled
RAW Offset(Or corrected)
image
Low power
Prep Ack
1. The detector is in low power state, user needs to send Cmd “Prep” to make the detector exit
to idle state which indicated by the acknowledge to Cmd “Prep”.
2. When the detector is in idle state, user can start the X-Ray any time.
3. When the X-Ray starts, the detector will sense the X-Ray automaticlly, the XWIN is configured
by parameter “Set Delay Time” on “Detector” page “Parameter” tab, the unit is “ms”, user
needs to make sure that the XWIN is larger than the X-Ray time.
4. After the XWIN is end, then the detector will start the acquisition flow.
5. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.