User's Manual

5. Operation Instructions For Image Acquisition
User Manual of Mars1717X 65
4.7.2. Work Flow(PrepAcq)
Select HWPostOffset HWGain HWDefect. If user need the raw image, please de-select all these
correction options.
Also, the software correction is supported.
IDLE Prepare
XWIN1
(fixed)
X-Ray
imaging
Cmd_PrepAcq Ack_Prohibit Ack_Enable
XWIN1
(fixed)
Offset
imaging
RAW Preview
RAW X-Ray image
(If HW-offset is disabled)
RAW offset(Or corrected)
image
1.
2. After receiving the Cmd_PrepAcq, it will start the prepare process, and send back the
3.
4. User needs to make sure the X-Ray ends within the XWIN.
5. The detector will send the images after the XWIN closed.
6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
4.7.3. Work Flow(Prep+Acq)
IDLE Prepare
XWIN2
(dynamic)
X-Ray
imaging
Cmd_Prep Ack_Prohibit Ack_Enable
XWIN2
(dynamic)
Offset
imaging
RAW Preview
RAW X-Ray image
(If HW-offset is disabled)
RAW offset(Or corrected)
image
Cmd_SingleAcq
X-Ray
1.