User's Manual

6. Regulatory Information
User Manual of Mars1717X 79
6.2
6.2.1 EMI Compliance Table
Emissions
Phenomenon Compliance Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class B
Professional healthcare facility environment
Harmonic distortion IEC 61000-3-2
Class A
Professional healthcare facility environment
Voltage fluctuations
and flicker
IEC 61000-3-3
Compliance
Professional healthcare facility environment
6.2.2 EMS Compliance Table
Enclosure Port
Phenomenon Basic EMC standard Immunity test levels
Professional healthcare facility environment
Electrostatic
Discharge
IEC 61000-4-2
8 kV contact
2kV, 4kV, 8kV, 15kV air
Radiated RF EM field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Near fields from RF
wireless
communications
equipment
IEC 61000-4-3
Refer to table fields from RF wireless
communications equipment
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, 5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745