User's Manual

Table Of Contents
Mars1717V Digital Flat Panel Detector User Manual
72 iRay Technology Co. Ltd..
5.1.2.2 Work flow
Click “Prep”
Panel preparation, Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X-ray at any time. However the longer time you wait. The worse
image would be. So please shoot once X-ray generator is ready.
After shooting, ClickAcquire
Wait for image uploaded
5.1.2.3 Timing Setting
To set a clear scenario for programming, see diagram below for details
1. Workstation receives “prep” request, send command “Clear” to panel.
2. Panel receives “clear” from workstation, starts clearing leakage of panel.
Meanwhile, panel send a message to workstation “Exposure Prohibited”.