5. Operation
iRay Technology Co. Ltd. 75
5.1.3.3 Timing Setting
To set a clear scenario for program, see diagram below for details
1. Workstation receives “prep” request and sends Clear” to panels.
2. Panel receives “clear” from Workstation, start clear operation. Meanwhile,
panel would send Exposure Prohibited” to Workstation.
3. Panel finishes”Clear” operation and send “Exposure Enable” to Workstation.
4. Workstation shows “Exposure Enable” on the iDetector’s message bar to tell
user shoot X-ray.
5. User triggers X-ray generator to initialize and do anode rotation to prepare for
X-ray shooting
6. X-ray generator finishes preparation and reminds users.
7. X-ray generator begins releasing X-ray
8. X-ray generator finishes X-ray shooting.
9. X-ray sensor in panel triggers panel to start image acquisition operation.
10. Panel completes image acquisition and begins to send data to Workstation.
11. Workstation starts receiving image data from panel.