iRay Technology Co. Ltd. 79
5.2 Correction and Calibration Template Generation
The correction and calibration should be performed after installation and it is
recommended to perform the new correction and calibration after any major
change on the system settings and hardware configuration. On the other hand, it is
also recommended to do the correction and calibration in each 6 months.
5.2.1 HW pre-offset Template Generation
Enter Acquire interface,
select HWPostOffset option
Enter Calibrate interface,
click UpdateHWPreOffset
button. Waiting until status
bar displayed: "Task
succeed:
HwGeneratePreOffsetTempl
ate”