User's Manual

SKR 40 0 0 DIRECT DIGITIZER User Manual
60
..................................................................................................................................................KONICA
MINOLTA, INC.
If the value meet the
expected value, click
“Accept”, then get the other 7
images.
If the value does not meet the
expected value, please do not
click the “Accept”, and
adjust the exposure
dose,thenclickthe“PREP
to get light image again
After getting 8 images, click
“Generate” to generate gain
template
Note:1pleaseusesoftwarepostoffsetmode.
2. Make sure your x ray dose is right, if your dose is out of the range, IDetector will remind you to
adjust the dose. Then you can click “start creating” and try again.
3. If users operate with two panels, SDK has a probability of quit automatically.
4.5. Image Check and upload
“OPEN” provides two features for image check and uploading. Local Image Check, Panel Image
Upload. Local Image Check defines function to check image saved in Workstation. Panel Image Upload
defines function to upload images stored in panel.
4.5.1. Local Image Check
Click Local File button in
“Local File” UI, choose the
specified file
Choose images stored in /