User's Manual

Table Of Contents
Appendix
A–40 AJ4310 | 2019-08-13
Correction and Calibration Management
Correction and Calibration Template Synchronization
The detector supports correction and calibration template storage. Templates in the
detector can be uploaded to a workstation, and templates in a workstation can be
downloaded to a detector.
1. After generating the offset, gain, and defect templates, select the templates and
select Download to FPD.
2. Select Download.