Appendix
A–42 AJ4310 | 2019-08-13
Correction and Calibration Management
The detector supports the following ways to do correction and calibration:
• In software correction and calibration, the workstation completes all correction and
calibration.
• In hardware correction and calibration, the panel completes all correction and
calibration.
1. User can set the calibration method on the Detector page.
Note:
The hardware-based calibration is on.