User's Manual

Table Of Contents
Safety and Regulatory Information
1–20 AJ4310 | 2019-08-13
Proximity Fields From RF Wireless Communications Equipment
Input AC Power Port
Test Frequency
(MHz)
Band (MHz)
Test Levels
Professional healthcare facility environment
385 380–390 Pulse modulation 18 Hz, 27 V/m
450 430–470 FM, ±5kHz deviation, 1 kHz sine, 28 V/m
710
704–787 Pulse modulation 217 Hz, 9 V/m745
780
810
800-–960 Pulse modulation 18 Hz, 28 V/m870
930
1720
1700–1990 Pulse modulation 217 Hz, 28 V/m1845
1970
2450 2400–2570 Pulse modulation 217 Hz, 28 V/m
5240
5100–5800 Pulse modulation 217 Hz, 9 V/m5500
5785
Emissions Test EMC Standard
Test Levels
Professional healthcare facility environment
Electrical fast transients/burst IEC 61000-4-4 ±2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1kV