User's Manual
Table Of Contents
- 1 Notices and Conventions
- Contents
- 1 Notices and Conventions
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings A-1
- Operating Modes A-8
- Software Installation A-12
- Set the Connection Mode A-13
- Shortcuts A-24
- Establish a Connection with the Detector A-25
- Configure the Detector A-26
- Correction and Calibration Template Generation A-29
- Image Check and Upload A-35
- Defect Template Check and Modification A-37
- Correction and Calibration Management A-40
- Update the Firmware A-44
- Publication History
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings
- Operating Modes
- Software Installation
- Set the Connection Mode
- Shortcuts
- Establish a Connection with the Detector
- Configure the Detector
- Correction and Calibration Template Generation
- Image Check and Upload
- Defect Template Check and Modification
- Correction and Calibration Management
- Update the Firmware
- Publication History
Installation
3–8 AJ4310 | 2019-08-13
Detector Battery Lock and Activation
To ensure the safety of the battery during transportation or storage, the battery can be
set to ship mode where it is locked and will not provide any voltage output.
Enter Ship Mode
Exit Ship Mode
Method Steps
Factory configuration
Web write in
1. Connect the adapter to the detector and
power it on.
2. Connect the other end of the adapter to
the PC.
3. Insert the battery into the detector.
4. Open a web browser and enter the
following address:
http://10.0.1.150/cgi-bin/ship-
mode.cgi?action=on
The battery is placed in ship mode.
Method Steps
Web exit
1. Connect the adapter to the detector and
power it on.
2. Connect the other end of the adapter to
the PC.
3. Insert the battery into the detector.
4. Open a web browser and enter the
following address:
http://10.0.1.150/cgi-bin/ship-
mode.cgi?action=off
The detector exits ship mode.