User's Manual

Table Of Contents
Appendix
A–8 AJ4310 | 2019-08-13
Operating Modes
The main purpose of the TRIMAX 35C is to acquire x-ray images. Most importantly, the
panel works in synchronization with the x-ray generator. TRIMAX 35C acquires x-ray
images in three ways: Software Mode, Inner2 Mode, and FreeSync Mode.
Software Mode
Block Diagram
Software mode is the basic way to acquire an x-ray image.
The workstation is a host device installed with iDetector and SDK. Installation describes
how to establish connections between the panel and workstation. In software mode, the
workstation does not control the x-ray generator. Users decide when to take x-rays.
Work Flow
1. Workstation receives prep request and sends clear command to the panel.
2. Panel receives clear from the workstation and starts clearing leakage from the panel.
Meanwhile, the panel sends an Exposure Prohibited message to the workstation.
3. Panel finishes clear and sends an Exposure Enable message to the workstation.
4. Workstation shows Exposure Enable on the iDetector’s message bar to tell the user
to take the x-ray now.
5. User triggers the x-ray generator to initialize and do an anode rotation to prepare for
taking x-rays.
6. X-ray generator finishes preparation for taking x-rays and reminds the user to take
the x-ray.
7. X-ray generator starts releasing the x-ray.
8. X-ray generator finishes taking the x-ray.
9. Workstation receives acquire request and sends Data Acquisition command to the
panel.