User's Manual
Table Of Contents
- 1 Notices and Conventions
- Contents
- 1 Notices and Conventions
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings A-1
- Operating Modes A-8
- Software Installation A-12
- Set the Connection Mode A-13
- Shortcuts A-24
- Establish a Connection with the Detector A-25
- Configure the Detector A-26
- Correction and Calibration Template Generation A-29
- Image Check and Upload A-35
- Defect Template Check and Modification A-37
- Correction and Calibration Management A-40
- Update the Firmware A-44
- Publication History
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings
- Operating Modes
- Software Installation
- Set the Connection Mode
- Shortcuts
- Establish a Connection with the Detector
- Configure the Detector
- Correction and Calibration Template Generation
- Image Check and Upload
- Defect Template Check and Modification
- Correction and Calibration Management
- Update the Firmware
- Publication History
Appendix
AJ4310 | 2019-08-13 A–29
Correction and Calibration Template Generation
The correction and calibration should be performed after installation and it is
recommended to perform the new correction and calibration after any major change to
the system settings and hardware configuration. It is also recommended to do the
correction and calibration every 6 months.
Pre-offset Template Generation
If panel is configured to do Pre-offset correction, Pre-offset Template is necessary.
1. Select Calibrate.
2. Select Start Generate Templates.
3. Select Create Offset.