Appendix
AJ4310 | 2019-08-13 A–31
2. Select Start.
3. Select PREP and start the exposure.
4. When completed, select Acquire to get the light image.
5. If the value meets the expected value, select Accept, and then acquire the other four
images.
If the value does not meet the expected value, do not click Accept. Adjust the
exposure dose, and then click PREP to acquire the light image again.