User's Manual
Table Of Contents
- 1 Notices and Conventions
- Contents
- 1 Notices and Conventions
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings A-1
- Operating Modes A-8
- Software Installation A-12
- Set the Connection Mode A-13
- Shortcuts A-24
- Establish a Connection with the Detector A-25
- Configure the Detector A-26
- Correction and Calibration Template Generation A-29
- Image Check and Upload A-35
- Defect Template Check and Modification A-37
- Correction and Calibration Management A-40
- Update the Firmware A-44
- Publication History
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- A Appendix
- Software and Settings
- Operating Modes
- Software Installation
- Set the Connection Mode
- Shortcuts
- Establish a Connection with the Detector
- Configure the Detector
- Correction and Calibration Template Generation
- Image Check and Upload
- Defect Template Check and Modification
- Correction and Calibration Management
- Update the Firmware
- Publication History
Appendix
AJ4310 | 2019-08-13 A–33
2. Select Start.
3. Select PREP and start the exposure.
4. When completed, select Acquire to get the light image.
5. If the value meets the expected value, select Accept, and then acquire the other
seven images.
If the value does not meet the expected value, do not select Acquire. Adjust the
exposure dose, then select PREP to acquire the light image again.