User's Manual

Operation
iRay Technology Co. Ltd. 32
synchronization with the X-ray generator, i.e., Software Mode and AED Mode.
4.1.1 Software Mode
4.1.1.1 Block Diagram
Software mode builds the first X-ray image acquisition step. Please see the figure below for
general features. Software mode is configured by selecting “prep” in Trigger mode and
“prepcapmode_acq2” in Prep capmode.
The workstation hosts the PC device installed with FDR SE Console or iDetector. Chapter 3.3
describes how to establish connection between detectors and the workstation. In Software
mode, the workstation does not control the X-ray generator; users decide when to take an X-ray.
4.1.1.2 Work flow
Detector clears until warning message changes
from “Exposure prohibittoExposure Enable”
Click “Acquire”
Shoot x ray in exposure window. But the longer
time user waits, the worse image would be
Wait for image uploaded