Instruction Manual

37
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ProTime® Microcoagulation System is intended for use in the electromagnetic environment specified
below. The customer or the user of the ProTime® Microcoagulation System should assure that it is used in
such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1 The ProTime
®
Microcoagulation System uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
The ProTime
®
Microcoagulation System is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.