g. Set the following Default Setup Group parameters:
Enter the desired Cor. Volume Unit parameter. Set Cor. Pulse Volume to the
same number of units.
For Output Pulse Spacing, enter the largest number over 285mS that works for
your application. Itron recommends a setting of 350 ms.
For Output Pulse Width, enter the largest number over 28 ms that works for your
application. Itron recommends a setting of 30 ms.
h. Click Send at the bottom of the page.