www.janitza.com Item no. 33.03.114 from firmware 1.1.x Doc. no. 1.040.015.4l Power Analyser Janitza electronics GmbH Vor dem Polstück 1 D-35633 Lahnau Support tel. 0049 6441 9642-22 Fax 0049 6441 9642-30 E-mail: info@janitza.com Internet: http://www.janitza.
UMG 96RM Table of Contents General Incoming goods inspection Scope of delivery of the basic device Available accessories Product description Intended use Characteristics of the basic device Measuring method GridVis network analysis software Connection options Assembly Installation Supply voltage Voltage metering Current measurement RS485 interface Digital outputs Operation Display mode Programming mode Parameters and measured values Configuration Applying the supply voltage Current and voltage transformers
UMG 96RM 3
UMG 96RM General Comments about the manual Copyright Your comments are welcome. If anything in this manual is unclear, please let us know and send us an e-mail at: info@janitza.com This manual is subject to the laws of copyright protection and may not be mechanically or electronically photocopied, reprinted, reproduced or otherwise reproduced or published in part or as a whole, without the legally binding, written consent of Janitza electronics GmbH, Vor dem Polstück 1, D 35633 Lahnau, Germany.
UMG 96RM Application notes Please read these operating instructions and all other publications that must be consulted in order to work with this product (particularly for installation, operation or maintenance). When using the device, the legal and safety regulations required for the respective application must also be observed. Please observe all safety regulations and warnings. Noncompliance with the instructions can lead to personal injury and/or damage to the product.
UMG 96RM About these operating instructions Incoming goods inspection These operating instructions are part of the product. • Read the operating instructions prior to using the device. • Keep the operating instructions at hand throughout the entire service life of the product and keep ready for referencing. • Hand over the operating instructions to each subsequent owner or user of the product.
UMG 96RM Scope of delivery of the basic device Quantity Item no. Designation 1 52.22.001 UMG 96RM 2 29.01.036 Mounting brackets. 1 33.03.113 Operating instructions. 1 51.00.116 CD with the following contents. - GridVis programming software - GridVis functional description 1 10.01.818 Screw terminal, pluggable, 2-pin (auxiliary energy) 1 10.01.828 Screw terminal, pluggable, 4-pin (voltage measurement) 1 10.01.820 Screw terminal, pluggable, 6-pin (current measurement) 1 10.01.
UMG 96RM Product description Intended use The UMG 96RM is provided for the measurement and calculation of electrical parameters such as voltage, current, power, energy, harmonics, etc. for building installations, to distributors, circuit breakers and busbar trunking systems. The UMG 96RM is suitable for installation in permanent, weatherproof switchboards. Conducting switchboards must be earthed. It can be mounted in any position.
UMG 96RM Characteristics of the basic device • General • Front panel-mounted with the dimensions 96x96 mm. • Connection via screw-type terminals. • LC display with backlighting. • Operation via 2 buttons. • 3 voltage measurements inputs (300V CATIII). • 3 current measurement inputs for current transformer. • RS485 interface (Modbus RTU, slave, to 115 kbps). • 2 digital outputs. • Working temperature range -10°C .. +55°C. • Storage of minimum and maximum values (without time stamp).
UMG 96RM Measuring method The UMG 96RM measures uninterrupted and calculates all root mean squares over a 9-period interval. The UMG 96RM measures the true root mean square (TRMS) of the voltages and currents applied to the measuring inputs. Operating concept There are several ways to program the UMG 96RM and retrieve measured values. • Directly on the device using two buttons. • Via the programming software of the GridVis. • Via the RS485 interface with the Modbus protocol.
UMG 96RM GridVis network analysis software Connection options The UMG 96RM can be programmed and read with the GridVis network analysis software which is part of the scope of delivery. For this, a PC must be connected to the RS485 interface of the UMG 96RM via a serial interface (RS485/Ethernet).
UMG 96RM Assembly Installation location Mounting The UMG 96RM is suitable for installation in permanent, weatherproof switchboards. Conducting switchboards must be earthed. The UMG 96RM is mounted on the switchboard by the side mounting brackets. These must be removed before using the device. Mounting is carried out by inserting and engaging the brackets. Installation position The UMG 96RM must be installed vertically in order to achieve sufficient ventilation.
UMG 96RM 13
UMG 96RM Installation Supply voltage L A supply voltage is required to operate the UMG 96RM. N The voltage supply is connected via plug-in terminals on the back of the device. Before applying the supply voltage, ensure that the voltage and frequency correspond with the details on the nameplate! Fuse Separator The supply voltage must be connected via a UL/IEC approved fuse (1 A, type C). Fig.
UMG 96RM c m Attention! The inputs for the supply voltage are dangerous to touch! • In building installations, the supply voltage must be provided with a disconnect switch or circuit breaker. • The disconnect switch must be attached near the device and must be easily accessible by the user. • The switch must be labelled as a separator for this device. • Voltages that exceed the permissible voltage range can destroy the device.
UMG 96RM Voltage metering The UMG 96RM can be used for voltage measurement in TN, TT and IT systems. Voltage measurement in the UMG 96RM is designed for the 300 V overvoltage category CATIII (4 kV rated pulse voltage). L1 In systems without a neutral, measured values that require a neutral refer to a calculated neutral. L1 277V/480V 50/60Hz L2 480V 50/60Hz L2 L3 L3 N PE Impedance V1 V2 V3 VN V2 V1 V3 VN Auxiliary energy Fig.
UMG 96RM Rated mains voltage Lists of the networks and their rated mains voltage in which the UMG 96RM can be used. Three-phase 4-wire systems with earthed neutral conductor. Unearthed three-phase, 3-wire systems. UL-N / UL-L UL-L 66 V/115 V 120 V/208 V 127 V/220 V 220 V/380 V 230 V/400 V 240 V/415 V 260 V/440 V 277 V/480 V 66 V 120 V 127 V 220 V 230 V 240 V 260 V 277 V 347 V 380 V 400 V 415 V 440 V 480 V Maximum rated voltage of the network Fig.
UMG 96RM Voltage measurement inputs The UMG 96RM has three voltage measurement inputs (V1, V2, V3). L1 L2 Overvoltage The voltage measurement inputs are suitable for measurement in networks in which overvoltages of overvoltage category 300V CATIII (4 kV rated pulse voltage) can occur. Frequency The UMG 96RM requires the mains frequency for the measurement and calculation of measured values. The UMG 96RM is suitable for measurements in the frequency range of 45 to 65 Hz. L3 N Fuse Separator Fig.
UMG 96RM When connecting the voltage measurement, the following must be observed: • A suitable separator must be provided in order to switch off the power to the UMG 96RM. • The separator must be placed near the UMG 96RM, marked for the user and easily accessible. • Use a fuse protected, UL/IEC approved 10A circuit breaker (type C) as an overcurrent protection device and separator.
UMG 96RM Connection diagram, voltage measurement • 3p 4w (addr. 509= 0), factory setting • 3p 4wu (addr. 509 = 1) L1 L1 L2 L2 L3 N L3 N V1 V2 V3 VN Fig. System with three-phase conductors and a neutral conductor. • 3p 4u (addr. 509 = 2) V1 L1 L2 L3 L3 V3 VN Fig. System with three-phase conductors and no neutral conductor. Measured values that require a neutral refer to a calculated neutral. 20 VN • 3p 2u (addr. 509 = 5) L2 V2 V3 Fig.
UMG 96RM • 1p 2w1 (addr. 509 = 4) • 2p 4w (addr. 509 = 3) L1 L1 L2 L3 N N V1 V2 V3 V1 VN Fig. Measured values derived from the V2 and V3 voltage measurement inputs are assumed to be zero and not calculated. • 1p 2w (addr. 509 = 6) V2 V3 VN Fig. System with uniform phase loading. The measured values for the V2 voltage measurement input are calculated. • 3p 1w (addr. 509 = 7) L1 L1 L2 L3 L2 L1 L2 L3 V1 V2 V3 VN Fig. TN-C system with single-phase, three-wire connection.
UMG 96RM Current measurement The UMG 96RM is designed for connecting current transformers with secondary currents of ../1A and ../5A. The factory set current transformer ratio is 5/5 A and may need to be adapted to the current transformers. It is not possible to perform a direct measurement without a current transformer with the UMG 96RM. Only AC currents (and not DC currents) can be measured. c m c 22 Attention! The current measurement inputs are dangerous to touch.
UMG 96RM Direction of the current The current direction can be individually corrected on the device or via the serial interfaces for each phase. In the case of incorrect connection, the current transformer does not need to be subsequently reconnected.
UMG 96RM Connection diagram, current measurement • 3p 4w (addr. 510= 0), factory setting • 3p 2i (addr. 510 = 1) L1 L1 L2 L2 L3 N L3 N I1 I2 I3 Fig. Measurement in a three-phase net-work with an unbalanced load. • 3p 2i0 (addr. 510 = 2) I1 • 3p 3w3 (addr. 510 = 3) L1 L2 L2 L3 L3 I2 I3 Fig. The measured values for the I2 current measurementinput are calculated. 24 I3 Fig. System with uniform phase loading. The measured values for the I2 current measurement input are measured.
UMG 96RM • 3p 3w (addr. 510 = 4) • 2p 4w (addr. 510 = 5) L1 L1 L2 L2 L3 N L3 N I1 I2 I3 Fig. System with uniform phase loading. The measured values for the I2 and I3 current measurement inputs are calculated. • 1p 2i (addr. 510 = 6) L1 I1 I2 I3 Fig. System with uniform phase loading. The measured values for the I2 current measurement input are calculated. • 1p 2w (addr. 510 = 7) L1 L2 N I1 I2 I3 Fig.
UMG 96RM Connection diagram, current measurement • 3p 1w (addr. 510 = 8) L1 L2 L3 L1 L2 L3 L1 L2 L3 I1 I2 I3 Fig. Three systems with uniform phase loading. The current measurement values of the phases of the respective system where are no CTs connected are calculated (I2/I3 resp. I1/I3 resp. I1/I2).
UMG 96RM Total current measurement If the current measurement takes place via two current transformers, the total transformer ratio of the current transformer must be programmed in the UMG 96RM. The UMG 96RM must then be set as follows: Primary current: 1000 A + 1000 A = 2000 A Secondary current: 5A UMG 96RM I1 1 Einspeisung 1 Supply 1 1P1 (K) (L) 1P2 1S1 (k) (l) 1S2 Verbraucher A Consumer A P1 1S1 1S2 2 P2 2S1 2S2 Example: The current measurement takes place via two current transformers.
UMG 96RM Ammeter If you want to measure the current not only with the UMG 96RM but also with the ammeter, the ammeter must be connected in series with the UMG 96RM. UMG 96RM I1 1 2 A Einspeisung Supply (k)S1 S2 (l) (K)P1 P2 (L) Verbraucher Consumer Fig. Current measurement with an additional ammeter (example).
UMG 96RM RS485 interface Terminating resistors The RS485 interface is designed with the UMG 96RM as a 2-pole plug contact and communicates via the Modbus RTU protocol (also see programming parameters). The cable is terminated with resistors (120 ohm 1/4 W) at the beginning and end of a segment. The UMG 96RM has no terminating resistors. Correct Incorrect A B RS485 interface, 2-pole plug contact Terminal block in the switch cabinet. Device with RS485 interface.
UMG 96RM Shielding Cable type A twisted and shielded cable must be provided for connections via the RS485 interface. The cable used must be suitable for an ambient temperature of at least 80 °C. • Ground the shields of all cables that run into the cabinet at the cabinet entry. • Connect the shield so it has a large contact area and conductively with a low-noise earth. • Mechanically trap the cable above the earthing clamp in order to avoid damage from cable movement.
UMG 96RM Bus structure • All devices are connected in a bus structure (line) and each device has its own address within the bus (also see programming parameters). • Up to 32 stations can be interconnected in one segment. • The cable is terminated with resistors (bus termination, 120 ohm 1/4 W) at the beginning and end of a segment. • If there are more than 32 stations, repeaters (line amplifiers) must be used in order to connect the individual segments.
UMG 96RM Digital outputs Digital outputs 1-2 Digital outputs 1-2 The digital outputs can switch DC and AC loads. The digital outputs are not short circuit protected. Connected cables longer than 30 m must be shielded. An external auxiliary voltage is required. The digital outputs can be used as pulse outputs. The digital outputs can be controlled via the Modbus. The digital outputs can output results from comparators.
UMG 96RM 33
UMG 96RM Operation The UMG 96RM is operated using buttons 1 and 2. Measured values and programming data appears on a liquid crystal display. A distinction is made between display mode and programming mode. The accidental changing of programming data is prevented by the entry of a password. Display mode In the display mode, you can scroll between the programmed measured value displays using buttons 1 and 2. All factory-set measured value displays listed in section 1 can be called up.
UMG 96RM Maximum value, HT/import Minimum value, NT/export Mean value Programming mode Sum measurement Phase conductorPhase conductor Password CT: Current transformer VT: Voltage transformer K1: Output 1 K2: Output 2 Button 2 Export Button 1 35
UMG 96RM Parameters and measured values Example of the parameter display All parameters necessary for operating the UMG 96RM, e.g. the current transformer data, and a selection of frequently required measured values are stored in the table. The contents of most addresses can be accessed via the serial interface and the buttons on the UMG 96RM. On the UMG 96RM display the value "001" is shown as the content of address "000".
UMG 96RM Button functions Display mode Programming mode Change mode Change mode Password simultaneous simultaneous Browse Browse short long short Measured values 1a Measured values 2a long Programming menu 1 Programming menu 2 long Measured values 2b short Programming menu 3 Programming Programming menu 1 Confirm selection Short: digit +1 Long: digit -1 (flashes) (flashes) Short: value x 10 (decimal to the right) Long: Value /10 (decimal to the left) 37
UMG 96RM Configuration Applying the supply voltage To configure the UMG 96RM, the supply voltage must be connected. c Attention! Supply voltages that do not correspond to the nameplate information can lead to device malfunction or destruction. The level of supply voltage for the UMG 96RM can be found on the nameplate. If no display appears, check the operating voltage to determine whether it is within the rated voltage range.
UMG 96RM C Current and voltage transformers The transformer ratios for each of the three current and voltage measurement inputs can be individually programmed in the GridVis software included in the scope of delivery. Only the transformer ratio of the respective group of current measurement inputs or voltage measurement inputs is adjustable on the device. Fig. Display for configuring the current and voltage transformers in the GridVis software.
UMG 96RM Programming current transformers Switching to programming mode • Simultaneously press buttons 1 and 2 in order to switch to programming mode. If a user password was programmed, the password request will appear with "000". The first digit of the user password flashes and can be changed with button 2. The next digit is selected by pressing button 2 and will begin flashing. If the correct combination was entered or if no user password was programmed, the device will enter programming mode.
UMG 96RM Programming voltage transformers • Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Use button 2 to switch to the voltage transformer setting. • Confirm the selection with button 1. • The first digit of the input area for the primary current starts flashing.
UMG 96RM Programming parameters Switching to programming mode • Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Use button 2 to switch to the voltage transformer setting. The first parameter of the parameter list is shown by repeatedly pressing button 2. Changing parameters • Confirm the selection with button 1. • The most recently selected address is displayed with the associated value.
UMG 96RM Device address (addr. 000) Mean value If several devices are connected to one another via the RS485 interface, a master device can only differentiate between these devices by means of their device addresses. Therefore, each device in a network must have a different device address. Addresses can be set in the range from 1 to 247. Mean values are formed over an adjustable for the current, voltage and power measured The mean values are identified with a bar the measured value.
UMG 96RM Averaging method Minimum and maximum values After the set averaging time, the exponential averaging method used achieves at least 95% of the measured value. All measured values are measured and calculated every 9 periods. Minimum and maximum values are determined for most of the measured values. The minimum value is the smallest measured value that has been determined since the last reset. The maximum value is the largest measured value that has been determined since the last clearance.
UMG 96RM Mains frequency (addr. 034) A voltage L-N of greater than 10V effective voltage must be present on at least one of the voltage measurement inputs in order that the mains frequency can be detected automatically. The mains frequency is then used to calculate the sampling rate for the current and voltage inputs. Adjustment range: 0, 45 .. 65 0= automatic frequency determination. The mains frequency is determined from the measurement voltage. 45..
UMG 96RM Energy meter Reading the active energy The UMG 96RM has energy meters for active energy, reactive energy and apparent energy.
UMG 96RM Harmonics Total Harmonic Distortion (THD) Harmonics are the integer multiple of a mains frequency. The voltage mains frequency for the UMG 96RM must be in the range between 45 and 65 Hz. The calculated voltage and current harmonics refer to this mains frequency. Harmonics up to 40x the mains frequency are recorded. THD is the ratio of the root mean square value of harmonics to the root mean square value of the mains frequency.
UMG 96RM Measured value relay Changeover time (addr. 039) All measured values are calculated every nine periods and can be recalled once per second on the measured value displays. Two methods are available for retrieving the measured value displays: Adjustment range: 0 .. 60 seconds If 0 seconds are set, no changeover takes place between the measured value displays selected for the measured value relay. The changeover time applies for all display change profiles.
UMG 96RM Display profile (addr. 037) Adjustment range: 0 .. 3 0 - Display profile 1, default setting. 1 - Display profile 2, default setting. 2 - Display profile 3, default setting. 3 - Customised display profile. C C The customised profiles (display change profile and display profile) can only be programmed via the GridVis software. Profile settings The profiles (display change profile and display profile) are clearly shown in the GridVis software included in the scope of delivery.
UMG 96RM User password (addr. 050) Clear energy meter (addr. 507) A user password can be programmed in order to impede any accidental change to programming data. A switch to the next programming menu can only be made after entering the correct user password. No user password is specified in the factory. In this case, the password menu is skipped and the current transformer menu is reached directly. The active, apparent and reactive energy meters can only be cleared together.
UMG 96RM Rotation field direction LCD contrast (addr. 035) The rotation field direction of the voltages and the frequency of phase L1 are shown on the display. The rotation field direction indicates the phase sequence in three-phase systems. Usually there is a "clockwise spinning rotation field". The phase sequence at the voltage measurement inputs is checked and displayed in the UMG 96RM.
UMG 96RM Time recording Operating hours meter The UMG 96RM records the operating hours and the total running time of each comparator The operating hours meter measures the time for which the UMG 96RM records and displays measured values. The time of operating hours is measured with a resolution of 0.1 h and is displayed in hours. The operating hours meter cannot be reset. • where the time of operating hours is measured with a resolution of 0.
UMG 96RM Serial number (addr. 754) The serial number shown by UMG 96RM has 6 digits and is part of the serial number displayed on the nameplate. The serial number cannot be changed. Serial number display Serial number information on the nameplate: XX00-0000 Software release (addr. 750) The software for UMG 96RM is continuously improved and expanded. The software version in the device is marked with a 3-digit number, the software release. The user cannot change the software release.
UMG 96RM Commissioning Applying the measured current Applying the supply voltage The UMG 96RM is designed for connecting ../1 A and ../5 A current transformers. Only AC currents and not DC currents can be measured via the current measurement inputs. Short circuit all current transformer outputs except for one. Compare the currents displayed on the UMG 96RM with the applied current. The current displayed by the UMG 96RM must match the input current, taking the current transformer ratio into consideration.
UMG 96RM Rotation field direction Checking the measurement Check the direction of the voltage rotation field on the measured value display of the UMG 96RM. Usually there is a "clockwise" spinning rotation field. If all voltage and current measurement inputs are correctly connected, the individual and sum power ratings are accurately calculated and displayed.
UMG 96RM Check the sum power ratings If all voltages, currents and power ratings for the respective phase conductor are correctly displayed, the sum power ratings measured by the UMG 96RM must also be correct. For confirmation, the sum power ratings measured by the UMG 96RM should be compared with the energy of the active and reactive power meters at the power feed.
UMG 96RM RS485 interface The data from the parameter and measured value list can be accessed via the MODBUS RTU protocol with CRC check to the RS485 interface. Address range: 1 .. 247 Factory default setting: 1 The device address is set to 1 and the baud rate is set to 115.2 kbps by default. C C The system does not support broadcast (addr. 0). The message length must not exceed 256 bytes.
UMG 96RM Example: Reading the L1-N voltage The L1-N voltage is stored in the measured value list under the address 19000. The L1-N voltage is stored in INT format. The UMG 96RM device address with the address = 01 is adopted here. The "query message" then appears as follows: Description Device address Function Start address Hi Start address Lo Disp. Values Hi Disp. Values Lo Error Check 58 Hex 01 03 4A 38 00 02 - Note UMG 96RM, address = 1 “Read Holding Reg.
UMG 96RM Digital outputs The UMG 96RM has 2 digital outputs.
UMG 96RM Digital outputs - status indicators The status of the switching outputs is represented in the UMG 96RM display by circular symbols. State of digital output 1 State of digital output 2 States of the digital output A current of <1 mA can flow. Digital output 1: Address 608 = 0 Digital output 2: Address 609 = 0 A current of <50 mA can flow.
UMG 96RM Pulse output Measured value selection (addr. 100, 101) Among other things, the digital outputs can also be used for the output of pulses to meter the energy consumption. After reaching a certain adjustable amount of energy, a pulse of defined length is applied to the output. Various adjustments must be made in order to use a digital output as a pulse output. Enter the power value here that is to be issued as an energy pulse. See Table 2.
UMG 96RM Pulse length (addr. 106) The pulse length applies for both pulse outputs and is permanently fixed via parameter address 106. Adjustment range: 1 .. 1000 1 = 10ms Default: 5 = 50ms Due to the minimum pulse length and minimum pulse pause, the values in the table are for the maximum number of pulses per hour. Pulse length Pulse pause Maximum pulses/hour The typical pulse length for S0 pulses is 30 ms.
UMG 96RM Pulse value (addr. 102, 104) The pulse value specifies how much energy (Wh or varh) should correspond to a pulse. The pulse value is determined by the maximum connected load and the maximum number of pulses per hour. If the pulse value is specified with a positive sign, pulses will only be issued if the measured value also has a positive sign. If the pulse value is specified with a negative sign, pulses will only be issued if the measured value also has a negative sign.
UMG 96RM Determining the pulse value External 230 V AC operating voltage Setting the pulse length Set the pulse length according to the requirements of the connected pulse receiver. For a pulse length of 30 ms, for example, the UMG 96RM can issue a maximum number of 60,000 pulses (see Table "Maximum Pulse Number") per hour.
UMG 96RM Limit value monitoring Two comparator groups are available for monitoring a limit value. Comparator group 1 is assigned to digital output 1 and comparator group 2 is assigned to digital output 2. State of digital output 1 UMG 96RM Addr. 608 =0 Source selection Comparator group 1 Digital output 1 Inverter Comparator B Logic Comparator A Result Addr. 616 0/1 Addr. 200 =0 Comparator C 0/1 Adr. 201=0 (not inverted) Adr.
UMG 96RM Example: Current monitoring in the neutral line If the current in the neutral line is greater than 100 A for 60 seconds, the digital output 1 should trip for at least 2 minutes. For the lead time of 60 seconds, any exceeding should be minimised. Address 112 = 60 seconds The following must be programmed: The operator for comparison between the measured value and the limit value. Address 113 = 0 (corresponds >=) 1. Comparator group 1 Select comparator group 1 for the limit value monitoring.
UMG 96RM Comparator Two comparator groups, each with 3 comparators, are available for monitoring limit values. The results from comparators A, B and C can be AND or OR linked. The linkage result from comparator group 1 can be assigned to digital output 1 and the linkage result from comparator group 2 is assigned to digital output 2. Comparator group 1 Comparator A Comparator B Comparator C Measured value (addr. 110) Limit value (addr. 108) Minimum turn-on time (addr. 111) Lead time (addr.
UMG 96RM • Measured value (addr. 110,116,122,129,135,141) The address of the measured value to be monitored is in the measured value. If measured value = 0, the comparator is inactive. • Limit value (addr. 108,114,120,127,133,139) Write the value in the limit that is to be compared with the measured value. • Minimum turn-on time (addr. 111,117,123,130,136,142) The linkage result (e.g. address 610) is maintained for the duration of the minimum turn-on time.
UMG 96RM Measured value Limit value Exceedance 2 seconds Lead time 2 seconds Minimum turn-on time Comparator result 69
UMG 96RM Service and maintenance The device is subject to various safety tests prior to delivery and is marked with a seal. If a device is opened, the safety tests must be repeated. A warranty is only given for unopened devices. Repair and calibration Firmware update If a firmware update needs to be implemented for your UMG 96RM, it can be implemented using the GridVis software (included in the delivery) via the Update Extras/ Device menu item.
UMG 96RM Error messages The UMG 96RM shows three different error messages on the display: Symbol for an error message - warnings, - serious error and - metering range exceedances. Error cause Description of the error If there are warnings and serious errors, the error message is indicated by the symbol "EEE" followed by an error number. Symbol for an error message Example of error message 911: The error number is composed of serious error 910 and internal error cause 0x01.
UMG 96RM Warnings Internal causes of the error Warnings are less serious errors and can be acknowledged with buttons 1 or 2. The measured values continue to be recorded and displayed. This error is redisplayed after each voltage recovery. The UMG 96RM can usually determine the cause of an internal error and then report it with the following error code. The device must be sent to the manufacturer for inspection. Error EEE 500 Description of the error The mains frequency cannot be determined.
UMG 96RM Metering range exceedance Metering range exceedances are displayed for as long as they are present and cannot be acknowledged. A metering range is exceeded if at least one of the three voltage or current measuring inputs is outside of its specified metering range. The phase in which the metering range exceedance occurred is indicated with the "up" arrow. The "V" and "A" symbols show whether the metering range exceedance occurred in the current or voltage circuit.
UMG 96RM Parameters of the metering range exceedance A continuative error description is stored encoded in the parameters of the metering range exceedance (addr.
UMG 96RM 75
UMG 96RM Procedure in case of error Possibility of error Cause Help No display External fuse for the power supply has tripped. Replace fuse. No current display Measurement voltage not connected. Connect measurement voltage. Measurement current not connected. Connect measurement current. Current measurement in the wrong phase. Check and correct the connection if necessary. Voltage transformer factor incorrectly programmed.
UMG 96RM Possibility of error Cause Help Ind./cap. phase shift The current circuit is assigned to the wrong voltage circuit. Check and correct the connection if necessary. Real power is too small or too large. The programmed current transformer ratio is incorrect. Reading and programming the current transformer ratio on the current transformer The current circuit is assigned to the wrong voltage circuit. Check and correct the connection if necessary.
UMG 96RM Technical data General Net weight 265 g Net weight (with attached connectors) 300 g Device dimensions approx. l = 42 mm, b = 97 mm, h = 100 mm Service life of the backlight 40,000 hours (50% of initial brightness) Transport and storage The following information applies for devices that are transported or stored in their original packaging.
UMG 96RM Supply voltage Installation overvoltage category 300V CAT II Protection of the power supply (fuse) 1 A, type C (approved by UL/IEC) Nominal range 95V - 240V (45..65Hz) or DC 100V - 300V Working area +-10% from the nominal range Power consumption max. 8.5VA / 3.5W Connection capacity of the terminals (power supply) Connectable conductor. Only one conductor may be connected per contact point! Single-wire, multi-wire, finely stranded conductor 0.2 - 2.
UMG 96RM Connection capacity of the terminals (outputs) Rigid/flexible 0.14 - 1.5 mm2, AWG 28-16 Flexible with ferrules without plastic sleeve 0.25 - 1.5 mm2 Flexible with ferrules with plastic sleeve 0.25 - 0.5 mm2 Tightening torque 0.22 - 0.
UMG 96RM Current measurement Metering range 0 .. 5 Arms (max. overload 7 Arms ) Crest factor 1.98 Resolution 0.1 mA (display 0.01 A) Overvoltage category 300 V CAT II Rated surge voltage 2 kV Power consumption approx. 0.2 VA (Ri=5 mOhm) Overload for 1 sec. 120 A (sinusoidal) Sampling rate 21.33 kHz (50 Hz), 25.6 kHz (60 Hz) per measuring channel Connection capacity of the terminals (voltage and current measurement) Connectable conductor.
UMG 96RM Parameters of functions Function Symbol Accuracy class Metering range Display range Total real power P 0.55) (IEC61557-12) 0 .. 5.4 kW 0 W .. 9999 GW * Total reactive power QA, Qv 1 (IEC61557-12) 0 .. 5.4 kvar 0 varh .. 9999 Gvar * Total apparent power SA, Sv 0.55) (IEC61557-12) 0 .. 5.4 kVA 0 VA .. 9999 GVA * Total active energy Ea 0.55) (IEC61557-12) 0 .. 5.4 kWh 0 Wh .. 9999 GWh * Total reactive energy ErA, ErV 1 (IEC61557-12) 0 .. 5.4 kvarh 0 varh ..
UMG 96RM Function Symbol Accuracy class Metering range Display range Current harmonics Ih Class 1 (IEC61000-4-7) up to 2.5 kHz 0 A .. 9999 kA THD of the current 3) THDi 1.0 up to 2.5 kHz 0 % .. 999 % THD of the current 4) THD-Ri - - - Mains signal voltage MSV - - - 1) Referred to amplitude. 2) Referred to phase and amplitude. (IEC61557-12) * The display returns to 0 W when the maximum total energy values are reached. 5) Accuracy class 0.5 with ../5 A transformer.
UMG 96RM Parameter and Modbus address list The following excerpt from the parameter list contains settings that are necessary for proper operation of the UMG 96RM, such as current transformers and device addresses. The values in the parameter list can be written and read. C In the excerpt, the measured value list files the measured and calculated measured values, output status data and recorded values so that they can be read.
UMG 96RM Address Adjustment Range Default 20 FLOAT RD/WR A Current transformer I2, sec. 22 FLOAT RD/WR V Voltage transformer V2, prim. 24 FLOAT RD/WR V Voltage transformer V2, sec. 26 FLOAT RD/WR A Current transformer I3, primary 28 FLOAT RD/WR A Current transformer I3, sec. 30 FLOAT RD/WR V Voltage transformer V3, prim. 32 FLOAT RD/WR V Voltage transformer V3, sec. 34 SHORT RD/WR Hz Frequency determination 0=Auto, 45 ..
UMG 96RM Address 50 Format RD/WR Unit Note Adjustment Range Default SHORT RD/WR - Password 0 .. 999 0 (no password) 0..32000 0 0..32000 0 -1000000..+1000000 0 -1000000..+1000000 0 1..1000 0,1 5 (=50 ms) 0 -1000000..+1000000 0 0..32000 0 0..32000 0..32000 0,1 0 0 0 -1000000..+1000000 0 0..32000 0 0..32000 0..
UMG 96RM Address Format RD/WR Unit Note 119 SHORT RD/WR Comparator 1B, Operator “>=”=0 “<”=1 120 FLOAT RD/WR Comparator 1C, Limit value 122 SHORT RD/WR Comparator 1C, Address of the measured value 123 SHORT RD/WR s Comparator 1C, Minimum turn-on time 124 SHORT RD/WR s Comparator 1C, Lead time 125 SHORT RD/WR Comparator 1C, Operator “>=”=0 “<”=1 126 SHORT RD/WR Result from comparator group 2; Link A, B, C (1=and, 0=or) 127 FLOAT RD/WR Comparator 2A, Limit value 129 SHORT
UMG 96RM Address Note Adjustment Range Default 142 SHORT RD/WR s 143 SHORT RD/WR s 144 SHORT RD/WR Comparator 2C, Minimum turn-on time Comparator 2C, lead time Comparator 2C, Operator “>=” = 0 “<” = 1 0..32000 0..32000 0,1 0 0 0 200 SHORT RD/WR 201 SHORT RD/WR 202 SHORT RD/WR 203 SHORT RD/WR - Select the source for Digital output 1 Digital output 1 inverter Select the source for Digital output 2 Digital output 2 inverter 0..4 *1 0..1 *2 0 0 0..4 *1 0..1 *2 0 0 -3..0..
UMG 96RM Address *1 *2 Format RD/WR Unit Note Adjustment Range 512 513 514 515 516 517 SHORT SHORT SHORT SHORT SHORT SHORT RD/WR RD/WR RD/WR RD/WR RD/WR RD/WR - Year Month Day Hour Minute Second 0..99 0..12 0..31 0..24 0..59 0..
UMG 96RM Table 2 - Modbus address list (frequently used measured values) C The addresses in the range from 0-999 listed in this document can be adjusted directly on the device. The address range over 1,000 can only be edited via Modbus! C A complete overview of the parameters and measured values as well as explanations regarding the selected measured values is filed in the document “Modbus Address List” on the CD or Internet.
UMG 96RM Modbus Address Address Above display Format RD/WR Unit Note 19036 19038 19040 19042 19044 19046 19048 19050 19052 19054 19056 19058 19060 19062 19064 19066 19068 19070 19072 19074 19076 19078 19080 19082 19084 19086 19088 19090 19092 19094 19096 876 878 880 882 820 822 824 800 - float float float float float float float float float float float float float float float float float float float float float float float float float float float float float float float RD RD RD RD RD RD RD RD RD R
UMG 96RM Modbus Address Address Above display Format RD/WR Unit Note 19098 19100 19102 19104 19106 19108 19110 19112 19114 19116 19118 19120 836 838 840 908 910 912 float float float float float float float float float float float float RD RD RD RD RD RD RD RD RD RD RD RD varh varh varh varh varh varh % % % % % % Reactive energy, inductive, L3 Reactive energy L1..L3, ind. Reactive energy, capacitive, L1 Reactive energy, capacitive, L2 Reactive energy, capacitive, L3 Reactive energy L1..L3, cap.
UMG 96RM 93
UMG 96RM Dimensional drawings All dimensions in mm. Rear view Side view 42 94 100 □ 91,5 □ 96 max.
UMG 96RM Bottom view Cutout dimensions 92+0,8 92+0,8 □ 91,5 97 95
UMG 96RM Overview of measured value displays A01 B01 C01 D01 Measured values L1-N voltage L2-N voltage L3-N voltage Mean values L1-N voltage L2-N voltage L3-N voltage Maximum values L1-N voltage L2-N voltage L3-N voltage Minimum values L1-N voltage L2-N voltage L3-N voltage A02 Measured values L1-L2 voltage L2-L3 voltage L3-L1 voltage B02 Mean values L1-L2 voltage L2-L3 voltage L3-L1 voltage C02 Maximum values L1-L2 voltage L2-L3 voltage L3-L1 voltage D02 Minimum values L1-L2 voltage L2-L3 vo
UMG 96RM A07 B07 C07 Measured values L1 apparent power L2 apparent power L3 apparent power Mean values L1 apparent power L2 apparent power L3 apparent power Maximum values L1 apparent power L2 apparent power L3 apparent power A08 Measured value Sum Apparent power B08 Mean value Sum Apparent power C08 Maximum value Sum Apparent power A09 B09 C09 Measured values L1 reactive power L2 reactive power L3 reactive power Mean values L1 reactive power L2 reactive power L3 reactive power Maximum val
UMG 96RM A13 B13 C13 Maximum value Distortion factor (THD) U L1 Maximum value Distortion factor (THD) U L2 Maximum value Distortion factor (THD) U L3 A14 Maximum value Distortion factor (THD) I L1 B14 Maximum value Distortion factor (THD) I L2 C14 Maximum value Distortion factor (THD) I L3 A15 Measured value L1 cos(phi) L2 cos(phi) L3 cos(phi) A16 Measured value Sum of cos(phi) B16 Mean value Sum of cos(phi) A17 Measured value Frequency L1 Rotation field display A18 Measured value Total a
UMG 96RM A19 B19 C19 Measured value (ind) Reactive energy Measured value Sum Reactive energy cap. Measured value Sum Reactive energy ind. A20 B20 Operating hours meter 1 Comparator 1 Total running time A21 B21 Measured value 1st. harmonic U L1 Measured value 3rd. harmonic U L1 A22 Measured value 1st. harmonic U L2 G20 ... Comparator 6 Total running time H21 ... Measured value 15th. harmonic U L1 B22 H22 Measured value 3rd. harmonic U L2 ... Measured value 15th. harmonic U L2 ...
UMG 96RM A25 B25 Measured value 1st. harmonic I L2 Measured value 3rd. harmonic I L2 A26 H25 ... Measured value 15th. harmonic I L3 H26 Measured value 3rd. harmonic I L3 A27 B27 Maximum value 1st. harmonic U L1 Maximum value 3rd. harmonic U L1 A28 Measured value 15th. harmonic I L2 B26 Measured value 1st. harmonic I L3 Maximum value 1st. harmonic U L2 ... H27 ... Maximum value 15th. harmonic U L1 B28 H28 Maximum value 3rd. harmonic U L2 ... Maximum value 15th. harmonic U L2 ...
UMG 96RM A31 B31 Maximum value 1st. harmonic I L2 Maximum value 3rd. harmonic I L2 A32 ... Maximum value 15th. harmonic I L2 ... Maximum value 15th. harmonic I L3 B32 Maximum value 1st. harmonic I L3 C H31 Maximum value 3rd. harmonic I L3 H32 Even and odd harmonics up to the 40th order can be called up via the GridVis software and can be viewed in the software. Marked menus are not displayed in the factory presetting.
UMG 96RM Declaration of conformity The UMG 96RM fulfills the following protection requirements: Directive 2004/108/EG in conjunction with DIN EN61326-1:2011 (IEC 61326-1:2010) as well as directive 2006/95/EG in conjunction with DIN EN 61010-1:2011 (IEC 61010-1:2011) and DIN EN 61010-2-030:2011 (IEC 61010-2-030:2011) Observed standards: Noise immunity DIN EN 61326-1:2006 (IEC 61326-1:2005) Electrical equipment for measurement, control and laboratory use – EMC requirements - part 1: General requiremen
UMG 96RM Connection example B 24V DC ~ K1 13 14 A K2 15 Digital outputs 16 17 B A UMG 96RM Basic device RS485 Supply voltage Voltage input Current input N/- L/+ V1 V2 V3 VN 1 3 4 5 6 2 S1 S2 3) 1) 2) 2) S1 S2 3) S1 S2 3) 2) L1 Consumer 1) L2 L3 N 2) 3) UL/IEC approved overcurrent protection device (1 A, type C) UL/IEC approved overcurrent protection device (10 A, type C) Jumpers (external) 230V/400V 50Hz 103
UMG 96RM Brief instructions Current transformer setting, primary current Changing the current transformer setting Display Programming mode Switch to programming mode: • Simultaneously press buttons 1 and 2 for approximately 1 second in order to switch to programming mode. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Confirm the selection with button 1. • The first digit of the input area for the primary current starts flashing.