User Manual
- 82 -
5.3 Troubleshooting
This product's states listed below are not always a sign of trouble.
Check for this product again before asking for repair.
Main unit state Check Items and Response
The main unit cannot be
turned on.
Nothing is displayed.
• Check if the battery pack is mounted.
→Mount a fully charged battery pack according to Section 3.1
Scan preparation.
• Check if the battery pack is exhausted.
→Mount a fully charged battery pack according to Section 3.1
Scan preparation.
• Check if the power plug of the AC adapter is connected.
→Connect the AC adapter correctly according to Section 3.1 Scan
preparation.
• Check if the battery pack is mounted while the output plug of the
AC adapter is connected to the AC input connector on this product.
→Disconnect the AC adapter according to according to Section
3.1 Scan preparation, and operate this product with the battery
pack.
The battery pack is
exhausted rapidly.
• Check if the battery pack is charged sufficiently.
→Charge the battery pack fully according to Section 3.10.2
Charging Device BC-3008 series.
• Has the battery pack itself reached the end of its life?
→ Exchange it for a new battery pack that is fully charged.
Scanning does not start
when the START key is
pressed.
• Are you at the parameter setting screen?
→Return to the scan screen following Section 2.3 Parameter
setting screen.
• Do you continuously press the START key?
→ Press the START key for more than one second. Without
pressing the key, continuously, the detection stops
automatically..
The B-mode display is
advanced when the
START key is simply
pressed.
• Check if time is set to the X-axis parameter (the scan mode).
→Set Distance according to Section 2.3.5 X-axis.
The detection is
automatically finished
during detection .
• Do you continuously press the START key?
→When you release the START key during detection, the
detection comes to an end after 8 seconds. (If the START key is
pressed again within 8 seconds, the detection continues.)