User Manual

- 82 -
5.3 Troubleshooting
This product's states listed below are not always a sign of trouble.
Check for this product again before asking for repair.
Main unit state Check Items and Response
The main unit cannot be
turned on.
Nothing is displayed.
Check if the battery pack is mounted.
Mount a fully charged battery pack according to Section 3.1
Scan preparation.
Check if the battery pack is exhausted.
Mount a fully charged battery pack according to Section 3.1
Scan preparation.
Check if the power plug of the AC adapter is connected.
Connect the AC adapter correctly according to Section 3.1 Scan
preparation.
Check if the battery pack is mounted while the output plug of the
AC adapter is connected to the AC input connector on this product.
Disconnect the AC adapter according to according to Section
3.1 Scan preparation, and operate this product with the battery
pack.
The battery pack is
exhausted rapidly.
Check if the battery pack is charged sufficiently.
Charge the battery pack fully according to Section 3.10.2
Charging Device BC-3008 series.
Has the battery pack itself reached the end of its life?
Exchange it for a new battery pack that is fully charged.
Scanning does not start
when the START key is
pressed.
Are you at the parameter setting screen?
Return to the scan screen following Section 2.3 Parameter
setting screen.
Do you continuously press the START key?
Press the START key for more than one second. Without
pressing the key, continuously, the detection stops
automatically..
The B-mode display is
advanced when the
START key is simply
pressed.
Check if time is set to the X-axis parameter (the scan mode).
Set Distance according to Section 2.3.5 X-axis.
The detection is
automatically finished
during detection .
Do you continuously press the START key?
When you release the START key during detection, the
detection comes to an end after 8 seconds. (If the START key is
pressed again within 8 seconds, the detection continues.)