Net-Ritef Continuity and Wiremap Tester Trace-Ritef Tone Generator and Amplified Tone Tracer User's Guide www.jdsu.
Voltage Probe Tracer/Power Button Volume Up/Down Indicator LED Power/Tone Type Switch RJ11
RJ11 LCD Screen Remote RJ11
Warning! Do not attach to AC power, this may damage the Net-Rite™ and the Trace-Rite™ generator and could cause a safety hazard for the user. Caution! Improperly crimped, damaged or un-crimped plugs can damage the jack on the NetRite™ and Trace-Rite™. Inspect plugs for proper termination and crimping before inserting into the tester. Contacts should always be recessed into the plastic grooves of the plug.
Features Net-Rite™ Features • Auto-on when sensing the remote or a short • Auto-off when the remote is disconnected to conserve battery life • Ergonomic shape, with soft edges • When not in use, the remote stores in the bottom of the main unit Trace-Rite™ Features • Self-powered Amplified Probe on the Tone Tracer turns on when tone sensor button is depressed and off when the button is released • Compact Tone Generator turns off automatically when stored in the bottom of the Tone Tracer to conserve battery
Instructions for Use To Wiremap a Cable 1) Using the RJ45 jacks, connect Net-Rite™ main unit and remote to the cable you want to test. ) An icon and the wiremap are always displayed. Net-Rite™ reports: a) PASS when the cable pins are properly connected per T568A/B and there are no other faults, or b) an error icon when it does not pass. Pins with errors flash an error message: Miswire, Split, Open, Short, or Reverse.
4) Adjust volume control on the side of the Trace-Rite™ to a comfortable level. If the signal is very loud when near the cable, adjust the volume to avoid overloading the Trace-Rite™. When overloaded, small increases or decreases in the signal at the tip cannot be heard. 5) Hold the tip of the Trace-Rite™ near the cable to be identified. The signal will be loudest on the wire or cable attached to the generator. Separating the wires or cables may help in isolating the correct cable.
Interpreting Wiremap Errors Miswire—One or more pins or pairs are not connected to the correct pins at the other end of the cable, i.e., one-to-one—pin 1 to pin 1, pin 2 to pin 2, etc. Split—A split pair is an error in the twisting of the wires together within the cable. The cables generally are made up of eight wires twisted together in 4 pairs. These 4 pairs are designated as pairs by the wiring standards and are intended to carry a signal and its return.
Reverse—A reverse (Rev) pair is a special case of a Miswire, and both icons will be flashing. The wires are connected to the correct pair of pins, but the two leads are reversed. Voltage—When voltage is detected on any of the pins, lightning bolt icons flash. Disconnect from the cable as soon as possible.
GREY=FLASHING Passing Shielded Cable 4-5 Pair Open 1-2 Short 3-4 & 4-5 Split Pair 2 & 3 Miswired
Battery Installation Net-Rite™ 1) Using a #1 Phillips screwdriver, unscrew and remove the battery door. ) Remove old cell batteries. Rapping the case against the palm of the hand will help dislodge the batteries. ) Install 4 new LR44 cell batteries, orienting the plus end of battery with the plus symbol on case. 4) Replace battery door and reinsert the screw. Do not over tighten.
Trace-Rite™ 1) Remove the screw at the back of the Tone Generator or Amplified Probe with a #1 Philips screwdriver and remove battery door. ) Remove old coin batteries. Rapping the case against the palm of the hand will help dislodge batteries. ) Install 4 new LR44 cell batteries being careful to orient plus end of battery with plus symbol on case. 4) Replace battery door and reinsert the screw. Do not over tighten.
To Replace the Probe Tip 1) Using a #1 Phillips screwdriver, remove the screw on the back of the Trace-Rite™ nearest to the probe. ) Pull the tip out of the top. ) Push the new tip into the top. 4) Replace the screw and tighten, being careful not to over tighten.
Models, options & accessories Description Ordering number Net-Rite continuity and wiremap tester and Trace-Rite tone generator with onboard probe kit KP260 Net-Rite continuity and wiremap tester TP250 Trace-Rite tone generator with on-board probe TT250 11
Net-Rite™ Specifications Physical Dimensions Size: 11.3 x 5 x 2.7 cm (4.5 x 2 x 1.1 inches) Weight: 100 grams (3.5 oz.), with battery and remote Environmental Operating temperature: 0 to 50˚C (32 to 122˚F) Storage temperature: –10 to 60˚C (14 to 140˚F Humidity: 10% to 90%, non-condensing Power Requirements Four (4) LR44 button cells (6 volts) Battery Life (6V alkaline battery, typical) times are for the full capacity of the battery Used continuously in one of the following modes Standby: 2.
Trace-Rite™ Specifications Physical Dimensions Size: 20.6 x 4.45 x 3.0 cm (8.1 x 1.75 x 1.
Customer Services This section provides a description of customer services available through JDSU (including returns policies and procedures) and warranty information. Customer Service (Standard Services) Customer Service accompanies the sale of every JDSU product.
manufactured equipment. This ensures products meet all published specifications, including any applicable product updates. Calibration — JDSU’s calibration methods are ISO approved and based on national standards. Factory Upgrades — Any unit returned for a hardware feature enhancement will also receive applicable product updates and will be thoroughly tested, ensuring peak performance of the complete feature set.
• A detailed description of the problem or service requested. • The name and telephone number of the person to contact regarding questions about the repair. • The return authorization (RA) number (US customers), or reference number (European Customers). If possible, return the equipment using the original shipping container and material.
Warranty Information JDSU guarantees that its products will be free of all defects in material and workmanship. This warranty extends for the period of 12 months for test instruments and 3 months for cables from date of manufacture or purchase (proof of purchase required). All product deemed defective under this warranty will be repaired or replaced at JDSU’s discretion. No further warranties either implied or expressed will apply, nor will responsibility for operation of this device be assumed by JDSU.
is reused, recycled, or disposed of in an environmentally friendly manner, and in compliance with all applicable national and international waste legislation. It is the responsibility of the equipment owner to return the equipment to JDSU for appropriate disposal. If the equipment was imported by a reseller whose name or logo is marked on the equipment, then the owner should return the equipment directly to the reseller.
www.jdsu.com/know Document Information Doc.