6.4 CREATING A NEW STANDARD CURVE
The user must firstly edit the calibration table before attempting to create a new standard curve.
Insert the cuvette containing the blank solution into the
reference position in the sample chamber and insert
the cuvette containing the first standard solution into
the sample position. Close the instrument lid and press
the start/stop button. Once the measurement is
complete the instrument will ask for the second
standard solution to be inserted. Repeat the
measurement procedure for each standard solution.
Once all standard solutions have been measured the
new standard curve and curve equation will be shown.
The standard curve can be viewed graphically by
pressing the function key below the show curve option
on the calibration table screen.