YUWELL® Finger Pulse Oximeter YX110 Please read the user manual carefully before using!(The picture is for reference only, please refer to the actual product.
Warning and Caution Safety: 1.Warning, Do not modify this equipment without authorization of manufacturer. 2.Warning, Keep away from the wet medical equipment such as drip or other similar liquid simulation as far as possible. 3.Warning, Check the sensor site every half an hour to ensure adequate blood circulation, intact skin, and appropriate sensor location. Otherwise, it may cause skin damage, compressive necrosis, or inaccurate measurement readings. 4.
electrosurgey. 14.Warning, Do not use the finger pulse oximeter in the presence of flammable anesthetics or other flammable substances, oxygen-enriched environments, or nitrous oxide to avoid the risk of explosion. 15.Caution,When using this product, you only need to fix it on your fingers. Excessive pressure on your fingers may cause skin damage. 16.
25.Warning, Properly apply and avoid using the oximeter in strong ambient light sources, fluorescent lights, infrared heat lamps, and direct sunlight to minimize interference that may result in unreadable or inaccurate readings. 26.Caution, The display may not read clearly when exposed to direct sunlight or bright light. 27.Caution,Do not use a functional tester to evaluate the accuracy of the Finger Pulse Oximeter. The functional tester shall only be used to check whether a unit is working properly. 28.
even endanger the patient's life. Therefore, it is very important to know oxyhemoglobin saturation of patient timely in clinical medical aspects. So that doctors can find problems in time. The YUWELL® Finger Pulse Oximeter features in small volume, low power consumption, convenient operation and portable.
locally applicable regulations , not with domestic waste 60529 Use-by date The device has no Alarm System Type BF applied part Humidity limitation Manufacturer Atmospheric pressure limitation Medical device Pulse rate (bpm:1/min) Temperature limit The Pulse Oxygen Saturation(%) Keep dry Rx only For prescription use This way up Fragile, handle with care Bluetooth Indication for Use Intended use: The YUWELL® Finger Pulse Oximeter is a non-invasive, non-sterile, reusable, spot checking device whi
Display Low battery Symbol of oxygen saturation Figure Oxygen saturation Symbol of pulse rate Pulse rate Pulse bar Figure 3 Technical Parameters 1. Display Type: LED SpO2 Display range: 0%~100%. Pulse Rate Display range: 25bpm~250bpm. 2. Power: Two AAA 1.5V alkaline batteries. 3. Working Current: Less than 40mA at rated voltage 3V. 4. .
Atmospheric pressure: 860hPa~1060hPa. 9. Operation mode: Continuous operation. 10.Device response time.(See Figure 4) 11.Peak wavelengths and light output power: Emission wavelength range 600nm-1000nm, radiation intensity is less than 15mW/sr (20mA). Information of wavelength range may be of especial use to clinical doctors. 12.Description of the effect on displayed and transmitted SpO2 and Pulse rate data value: 1)data averaging and other signal processing.
measured values of oxygen saturation and the reference value. The formula is as follows: Arms: accuracy. n: test sample quantity. SpO2i : measured value of pulse oxygen saturation during the first measurement using the finger pulse Oximeter. SRi: reference value of pulse oxygen saturation during the ith measurement using the carbon-monoxide-blood-gas analyzer. 4.
Product Properties 1. Operation of the product is simple and convenient. 2. The product is small in size, light in weight and portable. 3. The product can operate continuously for 17 hours with two brand new AAA batteries. (The operation time may vary due to the different performance of batteries) 4. Low voltage prompt will appear on the display when the battery voltage is lower than the minimum value of normal working voltage range. 5.
Please put or remove batteries in right order, or it will damage the device bracket. Please remove the battery if the oximeter is not used for long time. 3. Install as the figures show. (See Figure) Remove the battery from the product if it is not required for extended periods of time in order to avoid damage to the oximeter resulting from a leaking battery. Do not use the new batteries with the old ones. Alkaline battery of long service is recommended, and do not use rechargeable batteries.
1. Never immerse or soak the oximeter. 2. We recommend cleaning and disinfecting the oximeter before or after each use, or in accordance with the policies established by the hospital, to avoid long-term damage to the oximeter and avoid cross-infection. 3. Never use cleaning agents/disinfectants other than the recommended. 4. The sensor component is not cleaned and disinfected during testing. 5.
Note: Do not splash, dump any liquid into the oximeter and attachments, switch and connections, which may damage the oximeter. Problems Possible reason 1.Put finger incorrectly SpO2 or PR 2.Not used according to can not be recommended steps shown normaily 3.Nail polish or paste manicure Solution 1.Try again 2.Try some more times, if you can make sure about no problem exiting in the product, please go to a hospital timely for exact diagnosis. 3.Remove the nail polish or discharge manicure when measuring.
(e.g.deviation from the performance indicated in the instructions for use during IMMUNITY testing) and the finger pulse oximeter can recover from any interference within 30 seconds without affecting basic safety or performance without operator intervention.The following degradations, if associated with BASIC SAFETY or ESSENTIAL PERFORMANCE shall not be allowed: - component failures; - changes in programmable parameters or settings; - reset to default settings; and - change of operating.
Table 2 - For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer's declaration – Radiated Emission Emissions test IEC60601 test level Compliance level Result RF emissions CISPR 11 3m 40dB(μV/m) 30MHz-230MHz; 47dB(μV/m) 230MHz-1000MHz 3m 40dB(μV/m) 30MHz-230MHz; 47dB(μV/m) 230MHz-1000MHz Pass Table 3 - For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer's declaration – Electrostatic Discharge IMMUNITY test IEC60601 test level Compliance level Result Electrostatic discharge(ESD) I
frequency (MHz) (MHz) 385 380-390 TETRA 400 450 430-470 GMRS 460, FRS 460 704-787 LTE Band 13,17 710 745 780 810 870 930 1720 1845 1970 2450 (m) Test LEVEL (V/m) 0.3 27 0.3 28 0.3 9 Pulse modulationb) 2 18Hz 0.3 28 Pulse modulationb) 2 217Hz 0.3 28 Pulse modulationb) 2 217Hz 0.3 28 ) GSM 800/900, TETRA 800, 800-960 iDEN 820, CDMA 850, LTE Band 5 GSM 1800, CDMA 1900, GSM 1900 1700-19 DECT; 00 LTE Band 1,3,4,25;UMT S Bluetooth, WLAN, 2400-25 802.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. FCC Radiation Exposure Statement: The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure condition without restriction.
Product name: Finger pulse oximeter Model: YX110 S/N: MFG.DATE: JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO., LTD. Manufacturer Address:No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA www.yuwell.com Please reserve the warranty card carefully. Manufacturer:JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO.,LTD. Manufacturer Address:No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA For information about any YUWELL product, please contact:www.yuwell.