Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Foot Pedal Setup Z370584 Rev. C
5-30
Surgeon Settings
Advanced Control Pedal Calibration and Testing
Follow the steps below to calibrate and test your foot pedal.
NOTE: In addition to the procedure below, you can access the Advanced Control Pedal
Summary screen by pressing the foot pedal icon (shown below) near the top of the
screen.
1. On the Surgeons and Programs screen, tap .
2. Tap the Calibration button.
Figure 5-25 — Advanced Control Pedal Calibration
3. With your foot off the foot pedal, tap the Phase 1 button.
4. Put your foot on the foot pedal, press and hold the foot pedal all the way down and tap
the Phase 2 button.
5. Move the pedal to the far left and hold that position while you tap the Phase 3 button.
6. Move the pedal to the far right and hold that position while you tap the Phase 4 button.