Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Program Settings
Z370584 Rev. C Default Settings
6-13
Default Settings
Table 6.1 – 19 Gauge Program – Phaco Settings
Parameter Available
Settings
Default Settings
Advanced Infusion (Program based) On, Off Off
Function Phaco 1 Phaco 2 Phaco 3 Phaco 4
Submode name Phaco 1 Phaco 2 Phaco 3 Phaco 4
Pump selection Venturi
Peristaltic
Peristaltic Peristaltic Venturi Peristaltic
IV Pole Height 0–106cm or
0–42 inches
(Advanced
Infusion
disabled)
80 cm
31.5 inches
106 cm
42 inches
106 cm
42 inches
90 cm
35.5 inches
Vacuum control (Occlusion Mode/
CASE)
On, Off Occlusion Mode off
CASE off
Occlusion Mode off
CASE off
Occlusion Mode off
CASE on
Occlusion Mode off
CASE on
Min vacuum (Peristaltic and Venturi
pump)
Unoccluded (mmHg) 0 (P) and 0 (V) 200 (P), 50 (V) 200 (P), 50 (V) 25 (P), 0 (V)
Occluded 0 or 5 less than
occlusion threshold
setting
0 or 5 less than
occlusion threshold
setting
0 or 5 less than
occlusion threshold
setting
0 or 5 less than
occlusion threshold
setting
Max vacuum (peristaltic and Venturi)
Unoccluded (mmHg) 60 or 120 when
CASE on (P), 30 (V)
350 (P), 275 (V) 350 (P), 300 (V) 175 (P), 150 (V)
Occlusion threshold 40 240 240 120
CASE Parameters
CASE Upper threshold
(% of Unoccluded vac)
90 90 90 90
CASE up time (ms) 2000 2000 2000 2000
CASE vacuum
(% of Unoccluded vac)
80 80 80 80
CASE lower threshold
(% of Unoccluded vac)
60 60 60 60
CASE
one-touch
STD STD STD STD
Peristaltic pump ramp threshold (%)
/ Venturi Vacuum Ramp
(1 - Slowest - 5 Fastest)
50 / 1 30 /1 20 / 1 15 / 1S
Venturi pump On, Off Off Off On Off
Min flow/aspiration rate (peristaltic
and Venturi)
Unoccluded (cc/min) 10 10 10 10
Occluded 10 10 10 10
Max flow/aspiration rate (peristaltic)
Unoccluded (cc/mm) 20 36 36 32
Occluded 20 36 36 32
Fluidics mode type, aspiration Panel, Linear Panel Panel Panel Panel
Fluidics mode type, vacuum Panel, Linear Linear Linear Linear Linear
Min power
Unoccluded (%) 0 0 0 0
Occluded N/A N/A N/A N/A