Default Settings Z370584 Rev. C
6-14
Program Settings
Table 6.2 – 19 Gauge Program – IA, Vitrectomy, and Diathermy Settings
Max power
Unoccluded (%) 50 40 35 10
Power delivery Continuous Continuous Continuous Continuous
WHITESTAR
setting (duty
cycle)
On 6/3 (67%) On 8/4 (67%) On VWS (67%)
Quadrant 1 - 12/6
Quadrant 2 - 10/5
Quadrant 3 - 8/4
Quadrant 4 - 6/3
On 4/8 (33%)
Occluded mode (%) 50 40 35 10
Power delivery Continuous Multi-Burst Multi-Burst Continuous
WHITESTAR setting (duty cycle) On 6/3 (67%) On 8/4 (67%) On 8/4 (67%) On 4/8 (33%)
Power type, occluded and
unoccluded
Linear Unoccluded - Linear
Occluded - Panel
Unoccluded - Linear
Occluded - Panel
Linear
Short pulse rate, occluded and
unoccluded
1–14 pps 6 pps 6 pps 6 pps 6 pps
Long pulse rate, occluded and
unoccluded
1–6 pps 4 pps 4 pps 4 pps 4 pps
Pulse Shape Parameters
% Kick Low 5 5 5 5
% Kick High 5 5 5 5
Low Power Limit 0 0 0 0
High Power Limit 80 80 80 80
Initial Mode On, Off Phaco (first enabled
submode)
Surgical
Mode/
Submode
Aspiration Rate cc/
mm (Min/Max) for
Peristaltic Pump type
Vacuum mmHg (Min/Max),
Pump Type
Power % / Cut Rate
Cuts per Minute
(cpm) (Min/Max)
Pump/Vac Ramp IV Pole
Height (cm)
IA1 36 Panel 0 / 550
Linear, Venturi
NA 80% / 2 90
IA2 5 Panel 0 / 10
Linear, Peristaltic
NA 20% / 1 70
IA3 48 Panel 0 / 550
Linear, Venturi
*If change to Peristaltic, 600
Linear
NA 90% / 3 100
ICA
(Vitrectomy)
20 Panel 250 Panel, Venturi 50 / 2000 cpm
Panel
90% / 3 50
IAC
(Vitrectomy)
20 Panel 0 / 300
Linear, Venturi
50 / 500 cpm
Panel
90% / 3 60
DIA 1 NA NA 5 / 30
Linear
NA NA
DIA 2 NA NA 5 / 30
Burst
NA NA
Parameter Available
Settings
Default Settings