Operator's Manual

Table Of Contents
Default Settings Z370584 Rev. C
6-18
Program Settings
Table 6.6 – 21 Gauge Program – IA, Vitrectomy, and Diathermy Settings
Max power
Unoccluded (%) 50 40 35 20
Power delivery Continuous Continuous Continuous Continuous
WHITESTAR setting (duty cycle) On 6/3 (67%) On 8/4 (67%) On VWS (67%)
Quadrant 1 - 12/6
Quadrant 2 - 10/5
Quadrant 3 - 8/4
Quadrant 4 - 6/3
On 4/8 (33%)
Occluded mode (%) 50 40 40 20
Power delivery Continuous Multi-Burst Multi-Burst Continuous
WHITESTAR setting (duty cycle) On 6/3 (67%) On 8/4 (67%) On 8/4 (67%) On 4/8 (33%)
Power type, occluded and
unoccluded
Linear Unoccluded - Linear
Occluded - Panel
Unoccluded - Linear
Occluded - Panel
Linear
Short pulse rate, occluded and
unoccluded
1–14 pps 6 pps 6 pps 6 pps 6 pps
Long pulse rate, occluded and
unoccluded
1–6 pps 4 pps 4 pps 4 pps 4 pps
Pulse Shape Setting (WHITESTAR
ICE)
On, Off Off Off Off Off
Pulse Shape Parameters
% Kick Low 5 5 5 5
% Kick High 5 5 5 5
Low Power Limit 0 0 0 0
High Power Limit 80 80 80 80
Initial Mode On, Off Phaco (first enabled
submode)
Surgical
Mode/
Submode
Aspiration Rate cc/
mm (Min/Max) for
Peristaltic Pump type
Vacuum mmHg (Min/
Max), Pump Type
Power % / Cut Rate
cpm (Min/Max)
WHITESTAR
(ICE) Pulse
Shaping
Pump/
Vac Ramp
IV Pole
Height
(cm)
IA1 36 Panel 0 / 600
Linear, Venturi
NA NA 60% / 3 85
IA2 5 Panel 0 / 10
Linear, Peristaltic
NA NA 20% / 1 85
IA3 50 Panel 0 / 600
Linear, Venturi
NA NA 90% / 4 95
ICA
(Vitrectomy)
20 Panel 300 Panel, Venturi 50 / 2000 cpm
Panel
NA 75% / 3 50
IAC
(Vitrectomy)
20 Panel 0 / 350
Linear, Venturi
50 / 500 cpm
Panel
NA 75% / 3 60
DIA 1 NA NA 5 / 30
Linear
NA NA NA
DIA 2 NA NA 5 / 30
Burst
NA NA NA
Parameter Available
Settings
Default Settings