Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Program Settings
Z370584 Rev. C WHITESTAR Technology
6-25
WHITESTAR ICE Technology (Pulse Shaping)
The WHITESTAR ICE Technology was the next micro-pulse advance in phacoemulsification
technology, which combined modulated ultrasonic power (pulse shaping) with vacuum
control through the application of the Chamber Stabilization Environment (CASE).
This pulse shaping technology modifies the standard square wave pulse, by increasing the
amplitude of the two milliseconds of the on time “kick”, and then setting the remaining part
of the on time to the standard power setting. This is repeated for each on time period,
resulting in increased control and efficiency in phacoemulsification.
Figure 6-15 — WHITESTAR ICE Technology Pulse Shaping
To access the WHITESTAR ICE Technology settings:
1. Tap on the Settings screen.
2. Tap the WHITESTAR mode button.
3. Tap On for pulse shaping.
4. Tap Edit to access the pulse shaping parameter settings.
5. Tap Finished to close the screen. Tap Finished again to return to the main screen.
1. Kick Amplitude 3. Burst Width
2. 2 Millisecond Kick 4. Phaco Power Level