Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
FUSION Fluidics Phaco Z370584 Rev. C
6-32
Program Settings
Figure 6-22 — Phaco Surgical Screen with Occlusion Mode Power Settings
To adjust the occluded power delivery:
NOTE: Nonzero start parameters are the same for unoccluded and occluded settings. When
the current value reaches the occlusion threshold then the system uses the occlusion
settings. You cannot set the nonzero start value higher than the occlusion threshold.
1. Tap the Settings button.
2. Tap to change other power settings. A Power Settings screen opens.
3. Tap the buttons on the left of the Power Settings screen to select a power type.
4. Tap the On button to engage WHITESTAR Technology.
5. Tap Finished to close the screen.
1. VERITAS Swivel with ELLIPS
™
FX Technology
2. WHITESTAR Technology without Pulse Shaping