Error Messages, Troubleshooting, and Diagnostics
Z370584 Rev. C Troubleshooting
10-5
1. Disconnect the foot pedal cable from the console.
2. Attach the AC charger cable to the foot pedal cable.
3. Plug the AC charger into a power receptacle.
NOTE: When the green light is no longer blinking, the foot pedal batteries are charged.
Table 10.2 – Charging Options
* The AC Charger cannot be used for charging while the system is operating with VERITAS Foot Pedal, an ALP, or an ACP
in surgical modes.
Troubleshooting
General
The system does not come on when you turn the power switch to on
1. Turn the power switch off.
2. Confirm that you attached the power cord to the system back panel.
3. Confirm that you plugged the power cord into the electrical receptacle or another power
source.
4. Confirm that there is electrical power to the wall receptacle or to the power source.
5. Turn the system on.
6. If the system still does not come on, turn the power off. Check for bad fuses and replace
the fuse if necessary.
7. Contact JJSV for technical service.
The wireless VERITAS Remote Control does not pair
1. One VERITAS Vision System can only uniquely pair with one remote. If you try to pair
more than one remote at the same time, the pairing fails.
2. Only pair the remote with one console at a time. Do not press the pairing key-sequence
on multiple remote controls as this causes the pairing to fail.
The System Is: Advanced Linear Pedal
(ALP)
Advanced Control
Pedal (ACP)
VERITAS
Foot Pedal
Not attached to a power source
and the system is off
AC Charger AC Charger AC Charger
Attached to a power source and
the system is off
AC Charger AC Charger AC Charger
Attached to a power source and
the system is on
AC Charger or Wired Mode AC Charger or Wired Mode AC Charger or
Wired Mode
Attached to a power source and
the system is on and in-use
* Wired Mode * Wired Mode * Wired Mode