Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Error Messages, Troubleshooting, and Diagnostics
Z370584 Rev. C Troubleshooting
10-9
2. Check the flow rate setting.
3. Check the tubing fittings to the handpiece.
4. Check for kinks in the tubing.
5. Remove the handpiece and perform the test chamber test to ensure you have balanced
the handpiece.
6. Ensure you balanced irrigation and aspiration.
Phacoemulsification
No phacoemulsification
1. Ensure you selected the Phaco mode on the touchscreen.
2. Ensure the system is primed and tuned.
3. Check the foot pedal operation.
4. Ensure you properly connected the phaco handpiece cord to the phaco receptacle on the
front of the system.
5. Check the phaco power setting.
6. Ensure the phaco tip is tight on the handpiece.
7. Check for damage to the phaco tip.
8. If there is damage to the tip, replace the tip with a new tip and retune.
Poor or intermittent phacoemulsification
1. Check all the corrective steps above for “No phacoemulsification”.
2. Remove the phaco tip and then replace the tip. Ensure the tip is tight on the handpiece.
3. Check the phaco power delivery setting for both unoccluded and occluded (if applicable)
settings.
4. Tune the phaco handpiece.
Diathermy
No diathermy or poor diathermy
1. Ensure you selected the diathermy mode on the touchscreen.
2. Check the foot pedal operation.
3. Check the diathermy power setting.
4. Check the diathermy cord for a secure connection to the diathermy handpiece and to the
diathermy receptacles on the system.
5. Ensure the diathermy cord connections are dry.
6. Try to use diathermy starting at a low power setting and gradually increase the power.
7. Replace the diathermy cord.
8. Replace the diathermy handpiece.