Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Tubing Packs Z370584 Rev. C
3-6
System Components
VERITAS Advanced Infusion Tubing Pack
The VERITAS Advanced Infusion Tubing Pack (VRT-AI) (Figure 3-5) includes all the features
of the Advanced Fluidics Tubing Pack and allows for the activation of the Advanced Infusion
mode to increase irrigation pressure beyond what is supplied by gravity alone. The increase
in irrigation pressure is achieved by adding pressurized air (equivalent to 30 cm of bottle
height) into the balanced salt solution bottle. Compared to gravity-only tubing packs, the
Advanced Infusion feature allows greater irrigation pressure at the same IV pole height.
Figure 3-4 — VERITAS Advanced Infusion Tubing Pack
1. Spike (To Irrigation Bottle) 5. Irrigation Luer Lock Fitting
2. Manifold 6. Infusion Line
3. Drainage Bag 7. Infusion Connector (To Console Side Panel)
4. I/A Tubing Set (To Handpiece)
WARNING!
Bottles are the only irrigation source specified for use with the VERITAS Advanced
Infusion Tubing Pack. Using the VERITAS Advanced Infusion Tubing Pack with a bag
as the irrigation source could result in a lower irrigation pressure than what is
reported by the system.