Operator's Manual
Table Of Contents
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index
Foot Pedal Setup Z370584 Rev. C
5-10
Surgeon Settings
Figure 5-9 — Switches Screen – Single Linear Foot Pedal
The system displays the standard foot pedal switch assignment screen. The Single Linear
Foot Pedal Switches screen allows you to assign or change the action for each of the four
switches.
4. Select the major mode and then the applicable submode.
5. Tap the buttons under each switch name to assign or change the action for that switch.
The system displays a switch assignment dialog box with the text of the current switch
assignment shown in green.
6. In the PHACO Submode 1 mode, tap the Apply to All Phaco button to define the same
switches for the other 3 Phaco submodes.