Owner's Manual
Troubleshooting
24
25
23
Specifications
Ear/Object mode:
10℃~40℃(50℉~104℉), 15%~85%RH, non-condensing
00hPa ~ 1060hPaAtmospheric Pressure : 7
146.4*37.6*57mm
Approx.122grams(with batteries)
DC3V(2×AA battery)
Measuring range
Laboratory accuracy
Display resolution
0.1℃ or 0.1℉
Measure time
Approximately one second
Operating temperature
range:
Storage and transport
temperature range
-25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing
00hPa ~ 1060hPaAtmospheric Pressure : 7
Clinical accuracy
Dimension
Weight
Battery
Battery life
Approx. 800 readings
Measuring site Ear canal(Ear Mode)
Oral (This thermometer converts the ear temperature to
display its “oral equivalent.”)
Reference body site
Operation mode Ear mode(Adjust mode)
Ingress protecting rating
IP22
Shock
withstands drop of 3 feet
Expected service life
Three years
Ear mode: 32.0℃~43.0℃(89.6℉~109.4℉)
Object mode: 0℃~100℃(32℉~212℉)
Ear mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉)
at 15℃~35℃ (59.0℉~95.0℉) operating temperature range
±0.3℃ (0.5℉) for other measuring and operating
temperature range
Object mode: ±4% or ±2℃(4℉) whichever is greater
Troubleshooting
26
Contraindication
No contraindication
Battery Replacement
1. Replace battery when " appears in the buttom " of LCD
display. (See Figure 10)
2. Slide battery cover down as shown in Figure 11.
3. Remove battery and install 2 new AA alkaline batteries
as shown . in Figure 12
4. Slide battery cover back on.
Figure10
Figure11
Figure 12
0-1 year old:
Clinical bias:-0.05℃(-0.09℉);Clinical repeatability:
0.10℃(0.18℉);Limits of agreement:0.21℃(0.38℉)
1-5 years old:
Clinical bias:0.19℃(0.34℉);Clinical repeatability:
0.12℃(0.22℉);Limits of agreement:0.35℃(0.63℉)
Over 5 years old:
Clinical bias:0.16℃(0.29*℉);Clinical repeatability:
0.10℃(0.18℉);Limits of agreement:0.28℃(0.50℉)
In Ear mode:
Temperature taken is
higher than 43.0 ℃
(109.4℉).
In Object mode:
Temperature taken is
higher than 100 ℃
(212℉).
Problem
Solution
Read Temperature
Taking Hints thoroughly ,
then make sure the probe
cover and lens filter are
clean, then take a new
temperature measurement.
The thermometer works
properly.
Indicates battery is between
70% and 100% of usable
battery capacity.
Error message
Read Temperature Taking
Hints Thoroughly, then
check the integrity of the
probe cover and take a new
temperature measurement.
Remove probe cover in
object temperature
mode.
In object temperature
mode, measurement
with probe cover.
In Ear mode:
Temperature taken is
lower than 32.0 ℃
(89.6℉).
In Object mode:
Temperature taken is
lower than 0 ℃
(32℉).
Problem
The ambient temperature
is not within the range
between 10℃ and 40℃
(50℉~104℉).
The thermometer is
placed incorrectly
or unsteady.
The thermometer
showing a rapid ambient
temperature change.
The thermometer is not
functioning properly.
Error message
Measurement before
thermometer is ready
Solution
Place the
in a room for at least 30
minutes at room
temperature between
10℃ and 40℃
(50℉~104℉)
thermometer
Read Illustration For
Use thoroughly and
take a new temperature
measurement.
Allow the thermometer
to rest in a room for at
least 30 minutes at
room temperature:
between 10℃ and 40℃
(50℉~104℉)
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
Take a measurement
until or appears
on the display.
The thermometer works
properly.
Indicates battery is between
30% and 70% of usable
battery capacity.
It indicates that the power
is low, but you can
continue to measure.
Indicates battery is between
10% and 30% of usable battery
capacity.The thermometer will
take a proper measurement but
batteries must be replaced soon.
The thermometer could
not work due to low
battery.
Indicates battery is under 10%
of usable battery capacity.
Replace two new alkaline
batteries size AA.