Owner's Manual
28
30
28
29
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Electromagnetic environment-guidance
ComplianceEmissions test
The device uses RF energy only for its internal function.
Therefore, its emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Group 1
RF emissions CISPR 11
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Class BRF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
Electromagnetic Compatibility Information
Warning:
1. The Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.)
2. Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the [ME EQUIPMENT or ME SYSTEM], including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
Table 2
Electromagnetic Compatibility Information
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electromagnetic environm
entguidance
Compliance
level
IEC 60601 test level
IMMUNITY test
± 8kV
contact
± 2 kV ,
± 4 kV
± 8 kV,
± 15 kV air
± 8 kV
contact
± 2 kV ,± 4 kV
± 8 kV,
± 15 kV air
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
N/A
Surge
IEC 61000-4-5
± 1 kV
differential
mode
± 2 kV
common
mode
N/A
Voltage dips,
short interrupti-
ons and voltage
variations on p-
ower supply in-
put lines
IEC 61000-4-11
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycle
70% UT (30% dip in UT)
for 25 cycle
<5% UT (>95% dip in UT)
for 5 sec
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m;50Hz o r 6 0 H z
Power frequency magnetic fields
should be at levels charactertic of a
typical location in a typical comme-
rcial or hospital environment.
30A/m;50Hz
o r 6 0 H z
N/A
27
The device satisfies the EMC requirements of the international
standard IEC 60601-1-2. The requirements are satisfied under
the conditions described in the table below. The device is an
electrical medical product and is subject to special precautionary
measures with regard to EMC which must be published in the
instructions for use. Portable and mobile HF communications
equipment can affect the device. Use of the unit in conjunction
with non-approved accessories can affect the device negatively
and alter the electromagnetic compatibility. The device should
not be used directly adjacent to or between other
electrical equipment.
Electromagnetic Compatibility Information
This device complies with part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including
interference that may cause undesired operation.
The device has been evaluated to meet general RF exposure
requirement. The device can be used in portable exposure
condition without restriction.
FCC Information
静音要 改成短按 设置键, 长按切换 模式吗
记忆混 测