Operating Manual Cryo-Compact Circulators The HighTech Series CF31 CF41 1.951.4871-V2 19514871-V2.doc 06/13 JULABO GmbH 77960 Seelbach / Germany Tel. +49 (0) 7823 / 51-0 Fax +49 (0) 7823 / 24 91 info@julabo.de www.julabo.de 26.07.
Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been designed to help you gain an understanding of the operation and possible applications of our Cryo-Compact Circulators. For optimal utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.
Cryo-Compact Circulators TABLE OF CONTENTS Operating manual ............................................................................................................................5 1. Intended use .............................................................................................................................5 1.1. Description ...........................................................................................................................5 2.
8. Menu functions ............................................................................................................. 31 8.1. Configuration ..................................................................................................................... 32 8.2. Control ............................................................................................................................... 36 8.3. Start of a profile .......................................................................
Cryo-Compact Circulators Operating manual 1. Intended use JULABO Cryo-Compact Circulators have been designed for temperature application to specific fluids in a bath tank. The units feature pump connections for temperature control of external systems (loop circuit). JULABO circulators are not suitable for direct temperature control of foods, semi-luxury foods and tobacco, or pharmaceutical and medical products.
Operator responsibility – Safety instructions 2. Operator responsibility – Safety instructions The products of JULABO ensure safe operation when installed, operated, and maintained according to common safety regulations. This section explains the potential dangers that may arise when operating the circulator and also specifies the most important safety precautions to preclude these dangers as far as possible. The operator is responsible for the qualification of the personnel operating the units.
Cryo-Compact Circulators sheets). Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well ventilated areas. Only use recommended materials (bath fluids). Only use non-acid and non corroding materials. When using hazardous materials or materials that could become hazardous, the operator must affix the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible: 1 Warning label W00: Colors: yellow, black Danger area.
Operator responsibility – Safety instructions 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following European guidelines: Directive of the European Parliament and of the Council on the approximation of the laws of the Member States relating to machinery. Low voltage regulations with respect to legal harmonization of the member countries concerning electric devices for use within certain voltage limits.
Cryo-Compact Circulators 2.4.
Operator responsibility – Safety instructions Working temperature range Temperature stability Temperature selection via keypad remote control via personal computer Temperature indication: Resolution Pump pressure stage Temperature control Working temperature sensor Safety temperature sensor Absolute Temperature Calibration at internal control at external control Heater wattage (at 230 V) Heater wattage (at 115 V) Cooling capacity Medium ethanol Cooling compressor Refrigerant Cooling machine °C °C °C °C
Cryo-Compact Circulators Electrical connections: Computer interface RS232 / RS485 External sensor Pt100 Optional: (Order No.
Safety notes for the user 3. 3.1. Safety notes for the user Explanation of safety notes In addition to the safety warnings listed above, warnings are posted throughout the manual. These warnings are designated by an exclamation mark inside an equilateral triangle. “Warning of a dangerous situation (Attention! Please follow the documentation).” The danger is classified using a signal word. Read and follow these important instructions. Warning: Describes a possibly highly dangerous situation.
Cryo-Compact Circulators overflowing of the fluid. Prevent water from entering the hot bath oil. Do not drain the bath fluid while it is hot! Check the temperature of the bath fluid prior to draining (e.g., by switching the unit on for a short moment). Use suitable connecting tubing. Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding walls. Make sure that the tubing is securely attached. Regularly check the tubing for material defects (e.
Safety notes for the user The following questions should help to recognize possible dangers and to reduce the risks to a minimum. Are all tubes and electrical cables connected and layed? Note: sharp edges, hot surfaces in operation, moving machine parts, etc.
Cryo-Compact Circulators Operating instructions 4.
Operating controls and functional elements Mains power switch, illuminated 1 2 60 20 100 140 180 220 0 Adjustable excess temperature protection according to IEC 61010-2-010 °C 3 4 Setpoint1: IntAct : Power : Control : 150.00°C 148.53°C 80 % intern RS232 : ExtAct : Power : Control : 150.00°C 145.30°C 80 % extern LCD DIALOG-DISPLAY Line 1: Setpoint and origin of setpoint programming (Key / RS232 or RS485 / ext.
Cryo-Compact Circulators Display for the adjusted pump pressure stage in the -OFF- mode. Display for the effective pump pressure stage (rotation speed) after start. 4.3 Four stages, adjustable via the 4.4 5 button, in the menu >PUMP<.
Operating controls and functional elements Option: Electronic module Order No. 8 900 100 The Cryo-Compact Circulator automatically recognizes the connected electronic module.
Cryo-Compact Circulators 5. 5.1. Preparations Installation Place the unit on an even surface on a base made of nonflammable material. Cooling machine, pump motor and electronics produce intrinsic heat that is dissipated via the venting openings.! Never cover these openings! Be sure that the flow of ventilation can exit under the instrument.
Preparations Return flow safety device For this reason, shut-off valves can be integrated in the loop circuit. Order No. 8 970 456 8 970 457 Description Shut-off valve (suitable up to +90 °C) Shut-off valve (suitable up to +200 °C) Temperature application to external, closed systems SERIAL ALARM STAND-BY REG+ E-PROG ext Pt100 Unscrew the M16x1 collar nuts on the pump connectors with a 19 mm (3/4“) wrench and remove the sealing disks.
Cryo-Compact Circulators 5.3. Tubing Warning: Tubing: At high working temperatures, the tubing used for temperature control and for the cooling water supply represents a danger source. A damaged tubing line may allow a large amount of hot bath fluid to be pumped out within a short time. This may result in: Burning of skin Breathing difficulties due to hot atmosphere Safety instructions Use suitable connecting tubing. Make sure that the tubing is securely attached.
Preparations 5.4. Bath fluids Caution: Carefully read the material safety data sheet of the bath fluid used, particularly with regard to the fire point! If a bath fluid with a fire point of 65 °C is used, only supervised operation is possible. Water: The quality of water depends on local conditions. Due to the high concentration of lime, hard water is not suitable for temperature control because it leads to scale in the bath Ferrous water can cause corrosion, even on stainless steel.
Cryo-Compact Circulators 6. 6.1. Operating procedures Power connection Caution: Only connect the unit to a power socket with an earthing contact (PE – protective earth)! The power supply plug serves as a safe disconnecting device from the line and must always be easily accessible. Never operate the unit with a damaged mains power cable. Regularly check the mains power cables for damage.
Operating procedures Important: When using a bath fluid, the change in volume in case of change in temperature has to be respected. Fill in a little amount of bath fluid only so that the low level alarm is not triggered. Low level alarm is triggered at the following liquid level: CF31 75 to 80 mm below the tank rim CF41 80 to 85 mm below the tank rim When reaching the working temperature, check the liquid level. If the cooling coil is not completely covered with bath fluid, refill it. 6.3.
Cryo-Compact Circulators 7. 7.1. Manual operation Start - Stop 24.60 S 150.00 E ----Setpoint1: IntAct : Power : Control : 150.00°C 24.60°C 80 % intern Start: Press the start/stop key . The actual bath temperature is displayed on the VFD COMFORTDISPLAY. The LCD DIALOG-DISPLAY informs about adjustments and conditions on the Cryo-Compact Circulator (see example on the left). Stop: Press the start/stop key . The VFD COMFORT-DISPLAY indicates the message „OFF“. Autostart: see chapter 8.1.
Manual operation 7.3. Setting the temperatures Press the Factory setting: Setpoints >Setpoint1: Setpoint2: Sezpoint3: 20.00°C 37.00°C 70.00°C key to call up the menu for temperature selection. 3 different working temperatures are adjustable. Their values are freely selectable within the operating temperature range.
Cryo-Compact Circulators 7.4. Safety installations, warning functions Check the safety installations at least twice a year! (See page 14) 7.4.1. Excess temperature protection Settings for the excess temperature protection according to IEC 61010-2-010 and for the high and low temperature warning functions are made in a menu that is called up with the key . ESC Safety-values >Warn-Type: Warning OverTemp : 85.00°C SubTemp : -99.99°C SafeTemp : 50.00°C SafeSens : xx.
Manual operation Recommendation: Set the excess temperature protection >SafeTemp< at 5 to 10 °C above the working temperature setpoint. Warning: The excess temperature protection >SafeTemp< should be set at least 25 °C below the fire point of the bath fluid used.
Cryo-Compact Circulators Setpoint3 85.0°C OverTemp 87.0°C SubTemp 83.0°C The warning functions are only triggered when the actual bath temperature, after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3 seconds. SafeTemperature: Indicated is the cut-out value set with a screwdriver on the excess temperature protection device. Setting range: 20 °C ... 220 °C This safety installation is independent of the control circuit.
Manual operation 7.4.2. Early warning system, low level protection This low level protection is independent of the control circuit and is divided in two sections. 1. Switch in stage 1 recognizes a critical fluid level (patented) . An audible warning (interval tone) sounds and a message appears on the VFD COMFORT-DISPLAY XXXXX WARNING CODE 40 Press Niveau level warning Critical fluid level Replenish the bath tank with bath fluid Refill bath fluid! 2.
Cryo-Compact Circulators 8.
Menu functions 8.1. Configuration Main menu >Configuration Control Profile Start ESC ESC Configuration >Setpoint : keypad Autostart: off OFF-Mode : pump off Language : German Reset : no Variable : control Unit : °C 1. Use the up/down keys /RS232/ext.Pt100/eprog /on /pump on /Englich /yes /serial/eprog /°F to select the desired menu item. 2. Use the left/right keys to select the parameter (parameter blinks). 3. Press enter to store the new parameter.
Cryo-Compact Circulators eprog IntAct Power Control : : : : 50.0°C 24.7°C 80 % intern eprog – Can only be adjusted when an electronic module with analog connections is used (option). Setpoint setting via the analog interface REG+E-PROG connection with an external voltage or current source or a programmer. Important: Connect the external voltage or current source or a programmer to the circulator via the socket REG+E-PROG (see page 51).
Menu functions OFF-Mode Normally the circulating pump is switched via the start/stop signal. However, if circulation should be maintained also for the -OFF- condition, the parameter >pump on< needs to be set. Possible parameters: pump off (factory setting) pump on Language There are two options for the language of the LCD DIALOG-DISPLAY: German or English. Possible parameters: deutsch / englisch German / English Reset Use this to reset all values to factory setting (except date and time).
Cryo-Compact Circulators Warning: The working temperature range of the Cryo-Compact Circulator is determined during configuration. If set to >Control<, this range cannot be exceeded. If set to > serial < and > eprog <, heat or cold is applied to the bath without control. The permissible maximum temperature can be exceeded. The user has to take adequate precautions for temperature control.
Menu functions 8.2. Control The Cryo-Compact Circulator is conceived for internal and external temperature control. Switching is carried out in this submenu. Depending on what is set, only the respective set of parameters is indicated. Main menu >Control Profile Start Int. Programmer ESC internal ESC Control >ControlType: intern Selftuning : once /always/off Dynamic : aperio /normal Xp : 1.0K Tn : 100s 1 Tv : 5s Control >ControlType: extern Selftuning: once /always/off CoSpeed : 0.0 Xp : 0.
Cryo-Compact Circulators Sensor calibration of the external Pt100 sensor is carried out in the menu >ATCalibration<, with the > Status< set to >no< (see page 59). Suggested adjustments for external temperature control: AreaUp/ AreaDown and IntMax / IntMin see chapter >Limits< on page 56. External temperature control does not allow for setpoint setting via the socket „ext. Pt100“ (see page 32). Accessory: Pt100 external sensor Order No.
Menu functions Dynamic: This parameter affects the temperature pattern only in case of internal control. normal °C Possible parameters: temp. stability normal –Allows for reaching the setpoint faster, but overshooting of up to 5 % is possible. aperiodic setpoint aperiodic – (factory setting) Target temperature is attained without overshooting. t °C With both adjustments an adequate normal temperature stability is reached after approximately the same time. temp.
Cryo-Compact Circulators Setting range: internal / external 3 ...9999 s Resetting time >Tn< (Integral component) Compensation of the remaining control deviation due to proportional regulation. An insufficient resetting time may cause instabilities to occur. Excessive resetting time will unnecessarily prolong compensation of the control difference. Setting range: internal / external 0 ... 999 s Lead time >Tv< (Differential component) Setting range: 0.1 ... 99.
Menu functions 8.3. Start of a profile The start menu of the integrated programmer allows calling up and defined starting of one of six previously stored temperature profiles. The profiles are started manually or via the integrated timer. Main menu Control >Programmer Start Int.Programmer ESC ESC >Programmer Start >Start Profile 0 at Step 0 Runs 1 Profile End Stdby Start no 1 /SetpPG/Setp 1/Setp 2/Setp 3 /yes/time >hour.min Day.Mon Year Start 1. Use the keys Example: 11:15 11:15 11.
Cryo-Compact Circulators Profile End Stdby / SetpPG / Setp 1 / Setp 2 / Setp 3 This parameter is used to determine if the unit switches to the -OFFcondition at the end of a profile, or if temperature control is continued and the working temperature setpoint that is be used in such case. Stdby – If the end of the profile is reached, the Cryo-Compact Circulator switches to the -OFF- condition.
Menu functions Indication after successful start: The started programmer displays the actually calculated setpoint in line 1. The value changes until the final temperature of the section is reached (temperature ramp). 1st line: Examples: Setp. : IntAct: Prof. : Remain: Setp. : IntAct: Prof. : All : nd 125.8°C 124.7°C 4 Step: 12 00:03:45 125.8°C 124.
Cryo-Compact Circulators Deleting a complete section: Edit >Edit Profile: 1 **Step is deleted ** Setp: °C Time [hh:mm] Use the keys Use the keys Press the to move to line 2. to choose the desired section. key to delete the complete section. If the desired section is already chosen, the section is deleted even if the cursor > is in line 1, 3 or 4. Deletion is indicated in the respective line of the LCD display: ** Step is deleted ** The section that is currently processed cannot be deleted.
Menu functions 8.3.2. Interrupting a profile Setp. : 125.80°C IntAct: 124.70°C Prof. : 4 Step:12 *** pause *** to interrupt or continue a profile. The setpoint and time period set for the corresponding section are thus stopped at the values presently achieved. The instrument is put on hold and the message „pause“ flashes on the LCD DIALOG DISPLAY. A profile can be interrupted or restarted by an external emergency AK 2 1 Press the start/stop key 3 STAND-BY Setp. : xxx.xx°C IntAct: xxx.xx°C Prof.
Cryo-Compact Circulators 8.3.4. Termination of a profile Setp : IntAct: Prof. : Remain: 125.80°C 124.70°C 4 Step:12 00:03:45 Programmer Start >Start Profile at Step Runs Start x x 1 no A profile can be terminated by pressing . The programmer switches back to the Start menu. Press again to quit the menu or use the keys to remain in the Start menu. The start of another temperature profile can now be prepared if necessary.
Menu functions 8.4. Integrated programmer The integrated programmer allows any desired temperature program sequences to be realized. Such a temperature sequence is called profile. A profile consists of individual sections defined by duration (t:) and target temperature. Target temperature is the setpoint (T:), that is achieved at the end of a section. The programmer uses time and temperature difference values within a section to calculate a temperature ramp. Main menu >Int.
Cryo-Compact Circulators Int. Programmer >Edit Delete Print Set Clock Edit Important: Compile profiles Display sections Also a running profile can be changed. See „Edit after start“ page 42 Delete Delete sections Print Print a programmed profile Set clock Set the real time on the instrument Compile profiles: Example: Edit >Edit Profile: 1 Step: 6 Setp: 100.00°C Time [hh:mm] 01:05 A flashing segment indicates that a number needs to be entered. Under submenu „Edit Profile“ enter a profile number.
Menu functions Display sections (Step): Edit >Edit Profile: 1 Step: 3 Setp: 50.00°C Time [hh:mm] 00:02 1 Use the keys desired number to select the submenu „Step“, enter the and press enter or 2 If the submenu „Step“ is selected, section by section can be shown in ascending order by the key . Each previously set value is displayed. Example: Delete >Delete Prof. from Step to Step delete 4 8 34 yes Delete >Delete Prof.
Cryo-Compact Circulators 8.5. Analog inputs/outputs In order to use the analog inputs and outputs, the Cryo-Compact ALARM STAND-BY Circulator must be equipped with the electronic module available as option. Order No. 8 900 100 Electronic module REG+ E-PROG This submenu enables setting of the input and output values for the programmer input and the temperature recorder outputs of socket REG+E-PROG. The >STAND-BY< input and the >ALARM< output are configurable.
Menu functions Menu level 2 Inputs/Outputs >Chan.1 : ActInt Chan.2 : ActInt Chan.3 : Setpoint EPROG : Setpoint Ext.StBy : inactiv /activ AlarmOut : StdBy 1. Use the keys to select the desired (>) menu item in menu level 2 on the LCD DIALOG-DISPLAY. 2. Use the keys and to store the desired parameter or to enter menu level 3. An input window opens press enter for each menu item. 3. In the input window in menu level 3 select the menu item with . the keys 4.
Cryo-Compact Circulators 8.5.1. REG+E-PROG – temperature recorder outputs REG+E-PROG (11): Outputs of the connector CHANNEL 3 EPROG CHANNEL 2 1. First define the desired output value for channels 1 to 3: CHANNEL 1 REG+E-PROG ActInt internal actual temperature value (bath temperature) ActExt external actual temperature value (external sensor) Power periodic or intermittent heating or cooling Setpoint active setpoint temperature (Setpoint1, 2, 3,/ integr. programmer /external programmer) 2.
Menu functions 8.5.2. REG+E-PROG - programmer input Menu >Configuration< E-PROG - Input Configuration >Setpoint : eprog Keypad/RS232/extPt100 Autostart: off Standby : pump off Language : english Reset : no Variable : Control /serial/eprog Menu >Inputs/Outputs< Input EPROG >Value as: Flowrate Signal : Voltage L Value 0.0L H Value 50.0L Flowrate: xx.xL Setpoint/Variable (Liters/Minute) Setting needs to be carried out, if 1.
Cryo-Compact Circulators Eprog IntAct Power Control : : : : 50.0°C 24.7°C 80 % intern After returning the LCD display to standard display by pressing the temperature value adjusted and set on the external voltage or current source is displayed in line 1 (Example: Eprog : 50.00 C). This EPROG input enables the use of different voltage and current values as program parameters.
Menu functions 8.5.3. STAND-BY input Ext. StBy (10) (for external emergency switch-off) (Connector see page 67) Possible parameters: inactiv - standby input is ignored activ - standby input is active Activate the standby input: 1. Under menu item >Ext. StBy<, set the parameter to >activ<. 2. Connect an external contact ‚AK‘ (e.g. for emergency switch-off) or an alarm contact of the superordinated system.
Cryo-Compact Circulators 8.5.4. ALARM output Alarm output (9) (for external alarm signal) This socket is a potential-free change-over contact. With the adjustments in the menu >Inputs/Outputs< all operating conditions can be signaled without having to change the pin assignments.
Menu functions 8.6. Limits The menu >Limits< allows for setting the minimum and maximum values for all important setting ranges and capacity variables. Main menu >Limits Interface serial ATCalibration ESC ESC Limits >SetMax :300.00°C SetMin :-94.99°C HeatingMax: 100% CoolingMax: 0% IntMax :300.00°C IntMin :-94.
Cryo-Compact Circulators IntMax: and IntMin: Limit setting Setting range: see working temperature range page 9 The limits IntMax and IntMin are only valid for external control. IntMax and IntMin are used to limit the expected internal bath temperatures to any upper and lower values. The temperature controller cannot exceed these limits even if it would be required for reaching the temperature in the external system. Consequently the external setpoint may thus not be reached.
Menu functions 8.7. Serial interface For communication between Cryo-Compact Circulator and a PC or a superordinated process system the interface parameters of bath units must be identical. The adjustment is usually carried out one time only in the menu >Interface serial<.
Cryo-Compact Circulators 8.8. ATC Absolute Temperature Calibration, 3-point calibration ATC serves to compensate a temperature difference that might occur between Cryo-Compact Circulator and a defined measuring point in the bath tank because of physical properties. Main menu Interface serial >ATCalibration Pump ESC internal ESC ESC ATCalibration >Sensor : intern Status : no Type : 3-point TmpValue1: xx.xx°C CalValue1: xx.xx°C TmpValue2: xx.xx°C CalValue2: xx.xx°C TmpValue3: xx.xx°C CalValue3: xx.
Menu functions Type: A >1-point<, >2-point < or >3-point < calibration can be carried out. First geometrically define the location for calibration (measuring point TM), then determine the temperature values of of the calibration points. The calibration type also determines the number of pairs of values indicated on the LCD DIALOG-DISPLAY. TmpValue: 1 or 2 or 3 Defined temperature value of the calibration point. This value is automatically stored with >CalValue< and can be indicated for control purposes.
Cryo-Compact Circulators 1. Use the keys 2. With the keys to select the menu item. select the parameter (parameter blinks) or use the numeral keypad to enter the numbers (number segment blinks). 3. Press enter to store the new parameter or value. Example: Examples: TT = 80.00 °C TM = 79.73 °C TT = 120.00 °C TM = 119.51 °C TT = 160.00 °C TM = 159.34 °C 3-point calibration for internal control.
Menu functions 8.9. Setting the pump pressure Examples: Adjusted Effective The pressure of the circulating pump is adjustable in four grades. After setting, the VFD-Info-Display indicates the corresponding value. Adjustable pump capacity Illuminated display: stage 1 ... 4 for pump pressure Display for the adjusted pump pressure stage in the –OFF- mode. Display for the effective pump pressure stage (rotation speed) after start.
Cryo-Compact Circulators 9. Troubleshooting guide / Error messages XXXXX Alarm with a complete shutdown of the unit Whenever the microprocessor electronics registers a failure, a complete shutdown of the compressor, heater and circulating pump is performed. ALARM CODE 01 The alarm light „ “ illuminates and a continuous signal tone sounds. The VFD-COMFORT-DISPLAY indicates the cause for the alarm in form of a code.
Troubleshooting guide / Error messages ALARM CODE 06 Defect of the working or excess temperature sensor. Working temperature and excess temperature sensors report a temperature difference of more than 35 K. Other errors (I2C-BUS errors) Error in A/D converter Difference between working- and safetytemp. Bigger than permissible value. ALARM CODE 07 The internal I2C bus-system shows a failure. ALARM CODE 12 Error in A/Dconverter of the internal measuring system.
Cryo-Compact Circulators ALARM CODE 38 Ext. Pt100 sensor input without signal, but setpoint programming set to external Pt100. The early warning system for low level signals a critical fluid level. Replenish the bath tank with the bath fluid. Ext. Sensor input without signal but setpoint programming set to ext. Pt100. WARNING CODE 40 Niveau level warning Critical fluid level Replenish the bath tank with bath fluid Cancel the alarm state. Press the mains power switch off.
Electrical connections 10. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected to the plug housing. The unit ensures safe operation if connecting cables with a maximum length of 3 m are used. The use of longer cables does not affect proper performance of the unit, however external interferences may have a negative impact on safe operation.
Cryo-Compact Circulators 2 1 Socket for external Pt100 sensor Pt100 Shield 4 3 ext. Pt100 Plug 2 1 3 4 Pin assignment: Pin Signal 1 I+ 2 U+ 3 U4 I- Look on soldering side. The shield of the connecting cable is electrically connected to the plug housing and the sensor tube. Alarm output (for external alarm signal) This potential-free change-over contact is activated in case of an alarm when pins 2 and 3 are connected. Switching capacity Switching voltage Switching current ALARM max. max.
Remote control 11. Remote control 11.1. Setup for remote control 5 9 1 6 SERIAL R -OFFS I xxxx xxxx 1. Check the interface parameters for both interfaces (on Cryo-Compact Circulator and PC) and make sure they match. In the menu >Interface< set the menu item >Type< to >RS232< or >RS485<.( Serial interface see page 58) 2. In the menu >Configuration< set the menu item >Setpoint< to >RS232< or >RS485<. (>Configuration< see page 32) 3. Connect both units with an interface cable..
Cryo-Compact Circulators The commands are divided into in or out commands. in commands: asking for parameters to be displayed out commands: setting parameters The out commands are valid only in remote control mode. Examples: When the RS485 interface is used, the three-digit instrument address stands in front of each command. (example: address Ad32 = A032) Command to set the working temperature >Setpoint1< to 55.5 °C out_sp_00 55.5 A032_out_sp_00 55.
Remote control Command Parameter Response of Cryo-Compact Circulator out_mode_03 0 Set external programmer input to voltage. Voltage out_mode_03 1 0 V ... 10 V Set external programmer input to current. Current 0 mA ... 20 mA out_mode_04 0 Temperature control of internal bath. out_mode_04 1 External control with Pt100 sensor. out_mode_05 0 Stop the unit = R –OFF-. out_mode_05 1 Start the unit.
Cryo-Compact Circulators in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of Cryo-Compact Circulator version None Number of software version (V X.xx) status none Status message, error message (see page 72) in_pv_00 none Actual bath temperature. in_pv_01 none Heating power being used (%). in_pv_02 none Temperature value registered by the external Pt100 sensor. in_pv_03 none Temperature value registered by the safety sensor.
Remote control Command Parameter Response of Cryo-Compact Circulator in_par_11 none Tn control parameter of the cascade controller. in_par_12 none Tv control parameter of the cascade controller. in_par_13 none Adjusted maximum internal temperature of the cascade controller. in_par_14 none Adjusted minimum internal temperature of the cascade controller.
Cryo-Compact Circulators 11.5. Error messages Error messages Description -01 LOW LEVEL ALARM Low liquid level alarm -03 EXCESS TEMPERATURE WARNING High temperature warning -04 LOW TEMPERATURE WARNING Low temperature warning. -05 WORKING SENSOR ALARM Working temperature sensor short-circuited or interrupted. -06 SENSOR DIFFERENCE ALARM Sensor difference alarm. Working temperature and safety sensors report a temperature difference of more than 35 K.
JULABO Service – Online remote diagnosis 12. JULABO Service – Online remote diagnosis JULABO Cryo-Compact Circulator s of the HighTech series are equipped with a so-called black box. This box is implemented in the controller and records all significant data for the last 30 minutes. In case of a failure, this data can be read out from the unit by using special software. The respective program is available for free download from www.julabo.de \ EasyBlackBox.
Cryo-Compact Circulators 13. Installation of electronic module Caution: Always turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures, or before moving the unit. Configuration, installation, maintenance and repairs on the Cryo-Compact Circulator may only be carried out by qualified personnel. B ALARM STAND-BY REG+ E-PROG Turn off the unit with the mains switch and disconnect the power plug.
Cleaning / repairing the unit 14. Cleaning / repairing the unit Caution: Always turn off the unit and disconnect the mains cable from the power source before cleaning the unit. Prevent humidity from entering into the circulator. Service and repair work may be performed only by authorized electricians. To maintain the full cooling performance, clean the condenser (C) from time to time. OFF AUS SERIAL ALARM STAND-BY REG+ E-PROG ext Pt100 C T16A T16A T16A T16A 1.