9. Troubleshooting guide / Error messages
+
Whenever the microprocessor electronics registers a failure, a complete
shutdown of the compressor and circulating pump is performed. The alarm
light "
" illuminates and a continuous signal tone sounds.
The LED temperature display indicates the cause for the alarm in form of a
code.
Press enter
to quit the audible signal.
The recirculating cooler is operated without bath fluid, or the liquid level is
insufficient.
Replenish the bath tank with the bath fluid.
Tube breakage has occured (insufficient filling level due to excessive bath fluid
pumped out). Replace the tubing and replenish the bath tank with the bath
fluid.
Cable of the working temperature sensor interrupted or short-circuited.
Error in A/D converter
The return temperature is above the switch-off value of the high temperature
protection (85°C). Check dimensioning of application.
Use a stronger recirculating cooler if necessary.
The motor protection switch for the compressor motor is off. Set motor
protection switch to >1< . Check the fuses.
The pressure sensor of low pressure side (evaporation pressure) is faulty,
short-circuited or has a line interruption.
Have the repair done by a specialist.
The motor protection switch for the compressor motor is off. Set motor
protection switch to >1< . Check the fuses.
The pressure sensor of high pressure side is faulty, short-circuited or has a
line interruption.
Have the repair done by a specialist.
The suction gas temperature sensor is faulty, short-circuited or has a line
interruption.
Have the repair done by a specialist.
After eliminating the malfunction, press the mains power switch off and on
again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized service
station.