Remote control
60
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
version none Number of software version (V X.xx)
status none
Status message, error message (see page 61)
in_pv_00 none Actual bath temperature.
in_pv_01 none Heating power being used (%).
in_pv_02 none Temperature value registered by the external Pt100 sensor.
in_pv_03 none Temperature value registered by the safety sensor.
in_pv_04 none
Setpoint temperature of the excess temperature protection
in_sp_00 none Working temperature „SETPNT 1“
in_sp_01 none Working temperature „SETPNT 2“
in_sp_02 none Working temperature „SETPNT 3“
in_sp_03 none
High temperature warning limit „OVERTEMP“
in_sp_04 none
Low temperature warning limit „SUBTEMP“
in_sp_07 none Pump pressure stage
in_par_01 none Te - Time constant of the external bath.
in_par_02 none Si - Internal slope
in_par_03 none Ti - Time constant of the internal bath.
in_par_04 none CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05 none Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
in_par_06 none Xp control parameter of the internal controller.
in_par_07 none Tn control parameter of the internal controller.
in_par_08 none Tv control parameter of the internal controller.
in_par_09 none Xp control parameter of the cascade controller.
in_par_10 none Proportional portion of the cascade controller.
in_par_11 none Tn control parameter of the cascade controller.
in_par_12 none Tv control parameter of the cascade controller.
in_par_13 none Adjusted maximum internal temperature of the cascade
controller.
in_par_14 none Adjusted minimum internal temperature of the cascade
controller.
in_par_15 none Band limit (upper)
in_par_16 none Band limit (lower)