User Manual

Preparations
28
5.3. Temperature application to external systems
Caution: Securely attach all tubing to prevent slipping.
If the circulator is operated without external system, close the pump connector (22)
with the cap nut.
Temperature application to external, closed systems
T16A
T16A
ext
Pt100
SERIAL
ALARM
REG+
E-PROG
STAND-BY
21 22
The circulator is used for temperature application to external,
closed systems (loop circuit) with simultaneous temperature
application in the circulator bath.
Uns
crew the M16x1 collar nuts on the pump connectors with
a 19 mm (3/4“) wrench and remove the sealing disks. Using
the collar nuts, screw on the tubing connection fittings (for
tubing 8 mm or 12 mm in diameter) delivered with the unit
and tighten firmly. (Pressure pump: 22, suction pump: 21)
Push on the tubings, and secure with tube clamps.
Attach the tubings to the connectors of the external closed
system, e.g., an instrument with a pressure-resistant
temperature jacket or a temperature coil, and fasten with
tube clamps to prevent slipping.
Temperature application to external, open systems
S = Suction pump connection
D = Pressure pump connection
K = Float
H = Height adjustment
The circulator is equipped with both a pressure and suction pump
for external temperature application in open systems.
Differing flow rates of the pressure and suction pumps should be
compensated. To maintain a constant liquid level, the JULABO
„D+S“ Level Adapter is recommended for the external bath tank.
The flow rate of the pressure pump will be then regulated by a
built-in float device. The liquid level may be changed by a height
adjustment on the „D+S“ Level Adapter.
Accessory: „D+S“ Level Adapter Order No. 8 970 410
Important:
The liquid level should be equal in the internal and external
baths (absolute height).
If you take out samples (for example Erlenmeyer flasks) from
the external bath, turn the circulator off with the Start/Stop key.